6 Care and Maintenance | 6.3 Maintenance Work
ZEISS
6.3.3 Replacing the 12 V, 50 W Halogen lamp of the HAL 50 Halogen Illuminator
CAUTION
Burning hazard due to hot light sources
Light sources can become hot during processing.
4
Avoid touching the hot light source housing.
4
Let the light source cool down before touching it.
Prerequisite
ü
The microscope is switched off.
ü
The light source has cooled down for about 15 minutes.
Procedure
1. Remove the HAL 50 halogen light source
2
from the back of the stand
1
.
1
2
2. Put it down with the open side facing up.
3. Remove the used lamp
3
in upward direc-
tion.
3
4
5
5
4. Push the new lamp with its two cap pins gently and carefully into the socket
4
of the
HAL 50 halogen light source. Do not bend the cap pins.
5. Place the HAL 50 halogen light source with the connecting pins
5
into the back side of
the microscope and push until the lamp engages securely.
120
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001