OPERATION
ZEISS
Basic settings on Axio Observer 3, 3 materials stand
Axio Observer
106
431004-7244-001
12/2016
5.6
Basic settings on Axio Observer 3, 3 materials stand
Light Manager function on Axio Observer 3, 3 materials stand
The Light Manager function can be used both in reflected light (with HAL 100/microLED illuminator) and
transmitted light (with HAL 100 or microLED illuminator for transmitted light if the optional transmitted
light illumination mount is fitted).
The Light Manager on the Axio Observer 3, 3 materials performs the following functions:
−
Pressing the LM-Set button permanently saves the current brightness value of the objective in use, i.e.
the last value saved using the LM-set button is set again after switching off the microscope and
switching it on again.
−
Separate values are stored for reflected and transmitted light.
−
A light voltage is stored for each nosepiece position.
−
Dazzle protection function when changing the objective.
The default setting for the lamp voltage is 3 V for each objective position.
5.7
Basic settings on Axio Observer 5, 5 materials stand
After switching on, the instrument is initialized. The instrument status is shown on the LCD display.
LCD display during booting of the Axio Observer 5, 5 materials stand:
"Axio Observer 5 or 5 materials", the ZEISS logo and the status "Loading" are displayed while the
instrument is booting.