OPERATION
Carl Zeiss
Axio Imager operation and function controls (motorized version)
Axio Imager
82
430000-7344-001
M70-2-0020 e 06/2009
4.2
Axio Imager operation and function controls (motorized version)
In this section, those operation and function controls of the motorized version of the Axio Imager are
described that are different from the manual version. Controls not described here correspond to the
manual version.
As a great number of components are offered optionally, the specific design of your microscope may be
different from the illustrations below. So it is possible, for instance, to use also manually controlled
components (such as reflector turret, condensers, filter wheels) on the motorized stand. In this case,
however, the functionality of the microscope, particularly its operation via TFT display (touchscreen) will
be restricted.
Legend to Fig. 4-5:
1
Binocular phototube with motorized eyepiece shutter – Button for opening/closing the eyepiece shutter
2
Camera path deflection left, with interface 60N (built into stand) - Operation via TFT display (touchscreen)
3
Slot F for stop slider with centerable luminous-field diaphragm
4
Slot for 2-position filter wheels, discrete mot.: The filter wheels are not suitable for fluorescence applications.
5
Slot A for FL attenuator, discrete mot.
6
TFT display (touchscreen) for operation and adjustment of microscope (refer to Section 4.8)
7
Slot for 2-position filter wheels, discrete mot.
8
Motorized focusing drive, coarse/fine
9
Motorized objective nosepiece
10
Motorized condenser
11
Motorized reflector turret - Operation via TFT display (touchscreen)
12
Button for opening the luminous-field diaphragm
13
Button for closing the luminous-field diaphragm
14
Button for quick-action lowering of stage LOAD POSITION
15
Button for quick-action lifting of stage OPERATING POSITION
16
Button (setting depends on stand type)
17
Button (setting depends on stand type)
18
Button (setting depends on stand type)
19
Button for rotating nosepiece anticlockwise by one position (factory-set)
20
Button for rotating nosepiece clockwise by one position (factory-set)
21
Button for quick-action stage lifting OPERATING POSITION
22
Button for quick-action stage lowering LOAD POSITION
23
Button (not used by factory setting)
24
Button (not used by factory setting)
25
Button (not used by factory setting)
26
Button for turning reflector turret clockwise by one position (factory-set)
27
Button for turning reflector turret anticlockwise by one position (factory-set)
28
Button LM-Set
The assignment of keys 10 ... 23 can be changed individually on the TFT display (see
Section 4.8: Touchscreen -
Settings tab
).
Further motorized components (such as the motorized stop slider, motorized filter wheels,
motorized FL attenuator and the motorized tube lens turret) are to be operated via the buttons
directly arranged on the respective components.