6.6.1.3. Switching on the ion beam (FIB)
1
Go to the
FIB
tab of the
FIB Control
panel.
2
Ensure the
FIB Gun Pressure
is better than
5 x 10
-7
mbar.
CAUTION
Danger of arcing. Danger of damaging the ion source.
Before switching on the ion beam, ensure the FIB gun pressure is better than 5x10
-7
mbar.
3
Check the current status of the system
vacuum.
To be able to switch on the ion beam, the
system vacuum has to be
5 x 10
-5
mbar
or
better.
4
Click the
FIB
icon.