6.6.1.2. Adjusting tilt eucentricity
Before you can start imaging or milling, it might be necessary to adjust tilt eucentricity. By adjust-
ing the eucentricity, the specimen surface is moved into the tilting plane of the super-eucentric
stage. That is why the image does not shift out of the screen when the stage is tilted.
CAUTION
Danger of damaging objective lens or specimen if the specimen is too close to the objec-
tive lens.
Since the eucentricity is adjusted by using the M-axis, the working distance will be
changed during the eucentricity setup.
Ensure the working distance is 10 mm or more.
1
Open the
Panel Configuration
Bar
.
2
Double-click
FIB Daily Adjust
.
The
FIB Daily Adjust
panel opens.
3
Activate the
Crosshairs
checkbox to show the
crosshairs.
4
Center a characteristic feature in the middle of
the screen (i.e. in the middle of the crosshairs).
5
Click
Eucentric Axis
.
6
Click
Start
.
7
Follow the instructions in the wizard.
To re-centre the feature, use the Centre feature
function (
<Ctrl+Tab>
) or change X/Y.
To change the tilt degree of the stage:
a
Go to the
Stage
tab of the
SEM Control
panel.
b
In the
Go To T
(ilt) field, enter the
required degree.
How to
continue
Continue with switching on the ion beam (FIB).
Summary of Contents for AURIGA Compact Crossbeam
Page 1: ...AURIGA Compact Crossbeam workstation Instruction Manual ...
Page 50: ...RM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP PK aÉëÅêáéíáçå ìëíçãÉê ëÉêîáÅÉ ...
Page 54: ...RQ çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP RK fåëí ää íáçå ...
Page 150: ...NRM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Page 156: ...NRS çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NNK ÄÄêÉîá íáçåë ...
Page 160: ...NSM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...