6.6.4.3. Selecting deposition conditions
Preconditions:
•
Specimen has been moved to the coincidence point
•
The platinum precursor has been heated and outgassed
Procedure:
1
Switch on the electron beam:
a
Switch on the gun.
b
Switch on the EHT.
2
Go to the
FIB
tab of the
FIB Control
panel.
3
Ensure the
FIB Gun Pressure
is better than
5 x 10
-7
mbar.
CAUTION
Danger of arcing. Danger of damaging the ion source.
Before switching on the ion beam, ensure the FIB gun pressure is better than 5 x 10
-7
mbar.
4
Switch on the ion beam:
a
Go to the
FIB
tab of the
FIB Control
panel.
b
Click
On
.
5
Select
FIB
mode from the drop-down menu.