21
Changing the motor
AURA
®
handheld NIR - Rev. 2
EN_40_070_0030I | AURA
®
handheld NIR - Rev. 2 | Maintenance and Repair Manual
4
Dismantle the white reference assembly
Unscrew the two Torx screws
1
from the white reference mount
3
.
Use a TX 8 Torx screwdriver for this.
Remove the white reference assembly
2
.
Fig. 15
3
2
1
Dismantle the white reference assembly
5
Pull out the motor with cable
Unscrew the two Torx screws
2
from the motor holder
1
.
Use a TX 8 Torx screwdriver for this.
Pull the motor
3
out of the motor holder.
Fig. 16
2
3
1
Pull out the motor with its cable