What to do in the event of malfunctions
S100 / OPMI pico
Version 7.0
Page 218
G-30-1781-en
Lamp failure of xenon light source
Manual switching to the backup lamp
• Turn off the suspension system at the power switch before switching to
the backup lamp.
• Press the button (4) to open the lamp module.
The lamp module is slightly ejected.
• Pull out the lamp module (1) as far as it will go. Turn knob (2) through
180° until it snaps in place.
This moves the backup lamp into the illumination beam path.
• Push the lamp module (1) all the way back into the lamp housing.
• Turn on the suspension system at its power switch.
• In case the backup lamp is defective and an exchange of the lamp module
is not possible, please proceed with an external surgical lamp.
Failure of backup lamp
• If you work with the backup lamp and this lamp fails during a surgery, the
lamp module of the xenon light source must be exchanged, see Page 226.
CAUTION
Risk of burns!
In the event of overheating, the lamp switches off automatically. When the
lamp has cooled down, it automatically switches on again.
If the microscope is positioned unattended above the patient, the patient can
be burnt due to the produced heat.
• Immediately activate the backup lamp after a lamp failure or switch off
the light source completely.
• Check if the light source is switched off before leaving the device unat-
tended.
Summary of Contents for 000000-1080-349
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...