What to do in the event of malfunctions
Lamp failure of halogen light source
The lamp housing contains a backup lamp which can be manually swung into
the illumination beam path when the first lamp fails. Open flap (1) indicates
that the backup lamp is operative.
• Press button (2) to manually swing the backup lamp into the beam path.
Open flap (1) indicates that the backup lamp is operative.
• Replace the defective lamp at the first opportunity (see Page 224).
• If the backup lamp is defective and replacement is not possible, continue
surgery using an external OR illuminator.
Fig. 98: Swing in the
backup lamp
CAUTION
Risk of burns!
In the event of overheating, the lamp switches off automatically. When the
lamp has cooled down, it automatically switches on again.
If the microscope is positioned unattended above the patient, the patient can
be burnt due to the produced heat.
• Immediately activate the backup lamp after a lamp failure or switch off
the light source completely.
• Check if the light source is switched off before leaving the device unat-
tended.
1
2
Summary of Contents for 000000-1080-349
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...