S100 / OPMI pico
Preparations for use
Preparing the system for sterile use
Attach asepsis caps and drapes
Attaching asepsis caps
For sterile use, the system can be equipped with resterilizable products. The
asepsis sets available from ZEISS contain caps and handgrips which can be
sterilized in autoclaves. For detailed information on sterilization, please refer
to the enclosed instructions "Preparation of resterilizable products" for the re-
spective asepsis sets.
Attaching the drapes
To ensure sterile covering of the system, it is also possible to use the single-
use drapes.
When attaching the sterile drapes, make sure there is enough slack in the
drapes to allow for movement of the microscope carrier and surgical mi-
croscope. It is especially important that the drapes are completely loose
around the handgrips. The surgeon must be able to reliably operate the
controls through the drape.
CAUTION
Risk of infection caused by contamination!
• Only use the system with suitable, sterile accessories.
CAUTION
Risk of infection caused by contaminated asepsis accessories!
• Only use sterile asepsis accessories suitable for the device!
• Clean, disinfect and sterilize asepsis accessories before each use. This also
applies to the first use after delivery.
• Make sure that the operating staff is briefed about the sterile operation
of the device.
• Before each use of the surgical microscope, check whether the asepsis
caps are firmly attached to the surgical microscope.
NOTE
Overheating as a result of covering ventilation openings!
• Never cover any ventilation openings, e.g. with drapes! This may cause
the light source to overheat and lead to lamp failure.
Summary of Contents for 000000-1080-349
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...