18. May 2021
© Copyright 2021, PROCEQ SA
13
Units
The measured values (except Haze) can be displayed
either in Gloss Unit (GU) or as a percentage relative to the
selected measuring angle.
100%
@
20°
≡
2037 GU
100%
@
60°
≡
1000 GU
100%
@
85
° ≡
161 GU
Date / Time
Set up date, local time and the time in minutes after the
device will turn off when not used. If Auto Off Time is set
to “0” the device will not switch off automatically.
Info
- General information about the device will be shown
here e.g. when the factory calibration expires and a
recalibration from the manufacturer is recommended.
4.8.4 Modes
The ZGM 1130 supports three different measure modes.
Standard and Sequence mode are used to trigger one
measurement at a time. Whereas Scan mode measures
continuously with high speed as long as the measure button
is pressed. This enables to measure an average value over
a much larger area than just the measuring area itself.
In
Standard
mode,
only single measurements are taken
and the statistic values are reset when the mode has been changed or the same mode has
been chosen again.
In
Sequence
mode,
the number of single measurements which should be made and lead to
an average value can be predetermined. The statistic values are reset after this predefined
number of single measurements, when the mode has been changed or the same mode has
been chosen again.
In
Scan
mode,
the ZGM 1130 measures with high speed as long as the measuring button is
pressed. All single measurements completed during that time lead to one single average
value which will be added to the statistic display. Treated as one measurement. The statistic
values are reset when the mode has been changed or the same mode has been chosen
again.