The JMA-Optic contains lithium-ion batteries. When shipping the JMA-Optic,
transport instructions for Li-ion batteries must be adhered to. Additionally:
Labeling of the transport packaging according to valid regulations.
Storage at temperatures > 70°C can lead to premature aging of the
installed batteries.
The devices may not be operated in wet zones, damp areas, climatic
chambers, under-pressure-, overpressure-, or height-chambers.
The system is not intended for use in areas where there is a risk of
explosion, rooms used for medical purposes or in a flammable atmos-
phere (oxygen-enriched).
The devices should not be used near engines or transformers with a big
connected load, for example, or heavy current power lines, as electrical
or magnetic interference fields can distort the correct measurements
and/or render them impossible.
Do not operate the device in the immediate vicinity of heat sources
(radiators) or in direct sunlight, because IR radiation could affect the
accuracy of the measurements.
Seite 9/69
Page 9/47
JMA-Optic
2.2. Safety
2.2.1. Environmental conditions
The JMA-Optic is suitable for use in dry interior rooms, as can be found in clinics,
medical practices and laboratories.
Temperature:
0°C to +45°C
Air pressure:
700 - 1100 hPa
Relative humidity:
max. 85%, non condensing
2.2.2. Storage and transport
Storage and transport of the JMA-Optic should only be carried out in the original packaging
(hard case) provided by the manufacturer.
Temperature:
-20°C to +70°C
Relative humidity:
max. 95%, non condensing
Protect from moisture