The cleanliness and integrity of the optical components must be checked
before each use to ensure the accuracy of the system.
For face bow or UK sensor service, please always ship both components
together in the original case to your dealer or manufacturer.
Seite 31/69
Page 31/47
JMA-Optic
7 Functional checks, preparation, disposal
■
Regular maintenance of the system helps to prevent damage and guarantees its long-term
safety. All of the procedures described in these instructions for use concerning maintenance
and preparation of the system are to be carried out on a regular basis.
■
If the system or accessory parts show damage, they should be sent to the manufacturer for a
safety inspection. In the case of malfunctions and/or defects being suspected and/or ascertained,
the device has to be taken out of use immediately, labeled as
”Out of Use”
.
■
All maintenance and servicing work on the system or individual parts that are extended beyond
the activities described in the instructions for use may only be completed by the manufacturer
or authorized organizations.
■
Prior beginning the preparation, switch the measurement system off under all circumstances
and disconnect it completely from the power supply.
7.1. Regular checks and safety checks
■
To maintain the proper status of the electrical operating equipment, it is necessary to carry out
repeated inspections and technical safety checks (e.g. in Germany according to BGV A3 accident
prevention regulations as well as technical safety checks pursuant to the German Medical
Products Operator Ordinance).
■
For safety reasons, inspecting the proper state of all the connection cables, network cables, USB
power supply and power sockets is recommended before every use of the measurement system.
If parts are damaged it is necessary to replace them before using the measurement system again.
■
Immediate servicing measures are to be carried out by the manufacturer if:
a) any liquids/fluids entered the device.
b) the cables or plug connectors show any damage.
c) parts of the sensor technology are damaged.
d) covers are damaged or have come off.
e) a defect or fault is either suspected or ascertained.
■
If the name plate or other labeling (e.g. warning notices) on the machine are damaged or
unreadable, these are to be replaced.