background image

on the screen. 

The tester will save testing results automatically, including 7 testing results and 

1 average value each group, and total 100 groups can be stored. 

Note:  Please  store  the  data  by  pressing  the  “MEM”  key  before  the  average 

value is displayed in the current group test. 

5.7.2 Reading 

In saving status, press and hold the “MEM” button to enter the reading mode 
and the symbol “R” will be showed on the display. At the same time, the first 

impact  value  in  the  recently  saved  groups  will  be  displayed,  press  “up”  and 
“down” button to read the saved testing value. 

Then press (or long press) “MEM” to quit the storage function, and the symbol 
“R” disappears. 

 

 

 

 

 

 

 

 

5.7.3 Delete 

Press  “DEL”  button  in  reading  mode  to  delete  one  group  data.  Long  press 
“DEL” button to delete all stored values.   

5.7.4 Data Transmission 

The  data  in  stored  can  be  transferred  from  the  instrument  to  a  PC  via 

communication software, to  save, statistic and analyze the measurement. (See 

the help documentation in communication software for details) 

 

5.8 Blue tooth and print 

Optional

 

Press “Blue Tooth” button to connect with the Thermal Printer then the groups 

testing data will be printed. 

5.9 Calibration (Error-correction settings) 

Measuring error can be usually caused by test tip abrasion or changing another 

impact device. So it is necessary to correct the error to make the measurement 

accurate. 

Press “up” and “down” button at the same time after power on, then the tester 

will  enter  calibration  mode,  the  identification  of  calibration          appears. 

Impact the standard block continuously until get the average values, then press 
“up” and “down” button to adjust the value to the standard value, finally press 

the  store  button  to  finish  the  calibration,  the  tester  will  save  the  cal  data  and 

exit the calibration mode. 

If  you  want  to  give  up  the  calibration  suddenly,  please  press  the  “power  off” 

button directly in Calibration mode. 

6. Maintenance & servicing 

After  long  time  using,  the  impact  device  has  been  used  for  1000-2000  times, 

please use the cleaning brush to clean the tube and impact body. 

 

unscrew the support ring first, then take out the impact body; 

 

Spiral  the  nylon  brush  in  counter-clock  direction  into  the  bottom  of 

the tube and take it out for 5 times 

 

Install the impact body and support ring. 

 

Release the impact body after use. 

 

Any lubricant is absolutely prohibited inside the impact device. 

13                                                                                                                                                      14 

 

Summary of Contents for LM-100

Page 1: ...nts 8 4 1 2 Weight and Thickness Requirements 8 4 1 3 Surface Hardened Layer 8 4 1 4 Curved surface Requirement 8 4 2 Support of Test piece 9 5 Operation 9 5 1 Display 9 5 2 Keyboard 10 5 3 Turn on th...

Page 2: ...l voltage with the impact velocity The Leeb hardness values are calculated by the following formula HL 1000 Vb Va In Which HL Leeb hardness values Vb the voltage produced during the rebound of impact...

Page 3: ...n carbide Tungsten carbide Tungsten carbide Tungsten carbide Impact Device Diameter 20mm 20mm 20mm 30mm 20mm Impact Device Length 86 147 75mm 162mm 141mm 254mm 155mm Impact Device Weight 50g 80g 75g 2...

Page 4: ...uitcase 2 6 Optional accessories Impact device DC C DL D 15 G Standard Calibration Test Block Blue tooth module Thermal Printer Standard supporting ring Small Supporting ring Special supporting rings...

Page 5: ...s Mould cavity Heavy workpieces Failure analysis of pressure vessel turbine and other equipments Small test area The production line of Bearings and other parts Distinguish the material of the metal m...

Page 6: ...f a large weight and thickness may still cause deformation so reinforcement and support in the rear surface of the test point is necessary 5 Operation 5 1 Display 5 2 Keyboard 5 3 Turn on the instrume...

Page 7: ...irmly on the surface of the sample the impact direction should be vertical to the testing surface Press the release button on the upside of the impact device to test The sample and the impact device a...

Page 8: ...the Thermal Printer then the groups testing data will be printed 5 9 Calibration Error correction settings Measuring error can be usually caused by test tip abrasion or changing another impact device...

Reviews: