M A I N T E N A N C E   A N D   S T O R A G E

This section describes the proper procedures for care, maintenance and storage 

of the ProOBOD probe.  

Important:  Do not attempt to access the probe motor 

assembly or open the probe body.  Doing so will void 

any remaining warranty.  The probe body should only 

be opened by a YSI Authorized Service Center.  

UpDating instrUment anD prOBe firmware

The instrument and probe firmware can be updated via www.ysi.com/software.  

There you will find the new firmware files and instructions on how to update the 

instrument and/or probe. 

sensOr maintenance - DissOlveD Oxygen

CLEANING THE OBOD SENSOR CAP

The Sensor Cap should be kept clean since some types of fouling may consume 

or produce oxygen which could affect the dissolved oxygen measurements.  To 

clean the Sensor Cap, gently wipe away any fouling with a lens cleaning tissue 

that has been moistened with water.  Do not use a coarse towel or cloth or organic 

solvents to clean the Sensor Cap.  Using a coarse towel or organic solvent to clean 

the Sensor Cap may cause permanent damage to the cap.  For example, alcohol 

will dissolve the outer paint layer and other organic solvents will likely dissolve 

the dye in the cap.  If the sensor cap is damaged, it must be replaced.  

OBOD SENSOR CAP REPLACEMENT 

The frequency of sensor cap replacement is dependent on use.  For example, 

when the probe is powered on for approximately 4 hours per day, 5 days a week, 

the sensor cap will need replaced about once per year.  If the probe is powered 

on more than this, the sensor cap will need replaced more often.  To extend the 

sensor cap’s usable life, turn the instrument off when not in use.  For example, 

turn the instrument off over night, but not between readings.  The sensor cap 

will also need replaced if it is damaged or cracked.  See the 

Troubleshooting 

section of this manual for instructions on how to determine if the sensor cap 

needs replaced.  

When replacing a Sensor Cap, the Sensor Cap Coefficients must be manually 

updated in the ProODO instrument.  The instruction sheet shipped with the 

replacement  OBOD  sensor  cap  includes  the  calibration  coefficients  that  are 

specific to your Sensor Cap.  

u

IMPORTANT - Be sure to save the OBOD Sensor Cap instruction 

sheet in case you need to reload the calibration coefficients.

The replacement OBOD Sensor Cap is shipped in a humidified container and 

the package should not be opened until the cap is needed for use.  Once the 

sensor cap has been installed on the OBOD probe as described in this section, it 

is important to keep the sensor in a 100% humid environment.  Therefore, the 

ProOBOD sensor should be stored in either a BOD bottle that contains a small 

amount of clean water (approximately 40 ml) or in a BOD bottle  that is filled 

with clean water so the sensor cap is immersed in water, see

 

Sensor Storage for 

more information.   If the sensor dries out, refer to the Rehydration procedure in 

this manual for instructions on how to recondition the sensor cap. 

Refer to Figure 3 below when following the instructions for replacing the cap. 

 

Figure 3.

Caution:  Avoid touching the sensing end of the sensor 

cap during the following maintenance procedures. 

Remove the stir paddle from the probe by pulling it straight out.   

1. 

Remove  the  old  sensor  cap  from  the  probe  by  grasping  the  probe  body 

2. 

with one hand and then rotating the sensor cap counterclockwise until it is 

completely free.  Do not use any tools for this procedure. 

Carefully remove the o-ring and discard it.  Do not use any tools to remove 

3. 

the o-ring.

Clean the o-ring seat on the probe by first wiping off old o-ring lubricant 

4. 

with a dry lens cleaning tissue.  Next, clean away any build-up with a water-

moistened lens tissue and then dry with another lens tissue.  

Locate the o-ring supplied in the Sensor Cap replacement kit and install it 

5. 

on the probe.  Do not use any tools to replace the o-ring.  Be careful not to 

touch the clear optical DO sensor window.  Any fingerprints on the optical 

u

 

O-ring

 

OBOD Sensor 

Cap, shown 

here removed 

from probe

u

Summary of Contents for WE KNOW ProOBOD

Page 1: ...USER MANUAL ...

Page 2: ...ntenance and Storage 8 Updating Instrument and Probe Firmware 8 Sensor Maintenance Dissolved Oxygen 8 Sensor Maintenance Temperature 11 Sensor Storage 12 Troubleshooting 12 Dissolved Oxygen Readings 12 Help 13 Error Status messages 13 Restore Default Calibration Values 14 Probe Specifications 15 Accessories Parts List 16 Declaration of Conformity 17 Recycling 19 Contact Information 20 Ordering and...

Page 3: ...elected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by Failure to install operate or use the product in accorda...

Page 4: ...tional outlet adapters are also included The ProOBOD sensor measures dissolved oxygen in water using lifetime luminescence technology and uses a digital signal to send information between the instrument and probe The probe features a stirrer and motor to aid in sample movement to keep solids from settling at the bottom of the sample bottle and to speed up the response of the sensor Key advantages ...

Page 5: ...nstruction sheet is important because it contains the calibration coefficients for your sensor cap After using this sheet for general probe setup be sure to store it in a safe place in case you need to reload these calibration coefficients in the unlikely event that they are ever deleted from the instrument and probe Note A new cable probe assembly already has a sensor cap installed and the sensor...

Page 6: ...r cap needs replaced See the Sensor Maintenance Dissolved Oxygen section of this manual or the instruction sheet included with the replacement sensor cap for instructions on updating the Sensor Cap Coefficients When updating the Coefficients the sensor cap serial will be updated automatically based on your entries The Temperature Coefficient should not be modified unless instructed by YSI Technica...

Page 7: ...nd then press Cal to complete the calibration The message line at the bottom of the screen will display Calibrating Channel followed by Calibration Successful Press Esc to cancel the calibration i Temperature All ProOBOD probes have built in temperature sensors Temperature calibration is not required nor is it available To set the units press the Probe key on the ProODO instrument and select Displ...

Page 8: ...en replacing a Sensor Cap the Sensor Cap Coefficients must be manually updated in the ProODO instrument The instruction sheet shipped with the replacement OBOD sensor cap includes the calibration coefficients that are specific to your Sensor Cap u IMPORTANT Be sure to save the OBOD Sensor Cap instruction sheet in case you need to reload the calibration coefficients The replacement OBOD Sensor Cap ...

Page 9: ...Calibration Code has been entered correctly press enter to confirm Highlight K2 through KC in turn and use the 5 numeric entry screen to enter the corresponding new coefficient from the Calibration Code label as described in step 4 Press enter after each entry and then proceed to the next K selection After all the new coefficients have been entered highlight 6 Update Coefficients and press enter A...

Page 10: ... sensor cap and then recalibrate If you are still getting erroneous dissolved oxygen readings try replacing the sensor cap and then recalibrate If the erroneous readings continue contact YSI Technical Support to help determine the next step Inspecting the Sensor Cap for Damage Caution Avoid touching the sensing end of the sensor cap during the following maintenance procedures If DO measurements se...

Page 11: ... Values Occasionally the instrument may need to have the factory calibration default values restored To perform a factory reset press Calibrate highlight Restore Default Cal and press enter Select the parameter you wish to restore either DO or Barometer and press enter After selecting barometer or DO you will be asked to confirm the operation Highlight Yes and press enter P robe S pecifications Th...