background image

2

3

G E T T I N G   S T A R T E D

Throughout the manual, the term “probe” refers to the end of the cable where 

the sensor, stir motor, and stirrer are located.  The term “sensor” refers to the 

Optical Dissolved Oxygen sensing portion of the probe assembly and the term 

“sensor cap” refers to the removable sensing cap that is replaced about once per 

year (Figure 1).

Figure 1.

Unpacking the OBOD prOBe

Each ProOBOD probe and replacement OBOD sensor cap 

includes an instruction sheet with important information 

unique and specific to each individual sensing cap.  

Carefully  unpack  the  instrument  and  accessories  and  inspect  for  damage.  

Compare received parts with items on the packing list.  If any parts or materials 

are damaged or missing, contact YSI Customer Service at 800-897-4151 (+1 937 

767-7241)  or  the  authorized  YSI  distributor  from  whom  the  instrument  was 

purchased.

OBOD Sensor Cap, 

shown here removed 

from probe

Probe

 

 

 

Stir paddle

 

Optical DO sensor

 

 

Stirrer toggle switch, 

turns stir paddle 

on and off

Temperature sensor

i

Remove the ProOBOD cable/probe assembly from the shipping container and 

locate the instruction sheet that is included.  This instruction sheet is important 

because it contains the calibration coefficients for your sensor cap.   After using 

this sheet for general probe setup, be sure to store it in a safe place in case you 

need to reload these calibration coefficients in the unlikely event that they are 

ever deleted from the instrument and probe.  

Note - A new cable/probe assembly already has a 

sensor cap installed and the sensor cap coefficients 

are preloaded into the probe at the factory.  

Preparing the probe for the first time:

A new ProOBOD probe will have a clear protective cover placed over the 

1. 

sensor  cap  to  ensure  the  cap  remains  moist  during  shipment.    Carefully 

remove this clear cover from the end of the probe by pulling it straight off 

the sensor.  Save the clear cover along with the sponge in case you want to 

use it later for long term storage of the probe. 

Place the probe in a standard BOD bottle that contains a small amount of 

2. 

clean water (approximately 40 ml).  The sensor should not be immersed in 

water but rather in the air above the water.  The purpose is to create a 100% 

water-saturated air environment for the sensor which is ideal for calibration 

and storage when not in use.    

It is important to always keep your sensor in a moist 

environment so the sensor cap does not dry out.  (See 

Maintenance and Storage for more information.)  

u

i

Summary of Contents for WE KNOW ProOBOD

Page 1: ...USER MANUAL ...

Page 2: ...ntenance and Storage 8 Updating Instrument and Probe Firmware 8 Sensor Maintenance Dissolved Oxygen 8 Sensor Maintenance Temperature 11 Sensor Storage 12 Troubleshooting 12 Dissolved Oxygen Readings 12 Help 13 Error Status messages 13 Restore Default Calibration Values 14 Probe Specifications 15 Accessories Parts List 16 Declaration of Conformity 17 Recycling 19 Contact Information 20 Ordering and...

Page 3: ...elected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by Failure to install operate or use the product in accorda...

Page 4: ...tional outlet adapters are also included The ProOBOD sensor measures dissolved oxygen in water using lifetime luminescence technology and uses a digital signal to send information between the instrument and probe The probe features a stirrer and motor to aid in sample movement to keep solids from settling at the bottom of the sample bottle and to speed up the response of the sensor Key advantages ...

Page 5: ...nstruction sheet is important because it contains the calibration coefficients for your sensor cap After using this sheet for general probe setup be sure to store it in a safe place in case you need to reload these calibration coefficients in the unlikely event that they are ever deleted from the instrument and probe Note A new cable probe assembly already has a sensor cap installed and the sensor...

Page 6: ...r cap needs replaced See the Sensor Maintenance Dissolved Oxygen section of this manual or the instruction sheet included with the replacement sensor cap for instructions on updating the Sensor Cap Coefficients When updating the Coefficients the sensor cap serial will be updated automatically based on your entries The Temperature Coefficient should not be modified unless instructed by YSI Technica...

Page 7: ...nd then press Cal to complete the calibration The message line at the bottom of the screen will display Calibrating Channel followed by Calibration Successful Press Esc to cancel the calibration i Temperature All ProOBOD probes have built in temperature sensors Temperature calibration is not required nor is it available To set the units press the Probe key on the ProODO instrument and select Displ...

Page 8: ...en replacing a Sensor Cap the Sensor Cap Coefficients must be manually updated in the ProODO instrument The instruction sheet shipped with the replacement OBOD sensor cap includes the calibration coefficients that are specific to your Sensor Cap u IMPORTANT Be sure to save the OBOD Sensor Cap instruction sheet in case you need to reload the calibration coefficients The replacement OBOD Sensor Cap ...

Page 9: ...Calibration Code has been entered correctly press enter to confirm Highlight K2 through KC in turn and use the 5 numeric entry screen to enter the corresponding new coefficient from the Calibration Code label as described in step 4 Press enter after each entry and then proceed to the next K selection After all the new coefficients have been entered highlight 6 Update Coefficients and press enter A...

Page 10: ... sensor cap and then recalibrate If you are still getting erroneous dissolved oxygen readings try replacing the sensor cap and then recalibrate If the erroneous readings continue contact YSI Technical Support to help determine the next step Inspecting the Sensor Cap for Damage Caution Avoid touching the sensing end of the sensor cap during the following maintenance procedures If DO measurements se...

Page 11: ... Values Occasionally the instrument may need to have the factory calibration default values restored To perform a factory reset press Calibrate highlight Restore Default Cal and press enter Select the parameter you wish to restore either DO or Barometer and press enter After selecting barometer or DO you will be asked to confirm the operation Highlight Yes and press enter P robe S pecifications Th...

Reviews: