OPERATIONS MANUAL
ba76140e02 10/2017
4320(W)
CONDUCTIVITY CELL FOR ULTRAPURE WATER
Page 1: ...OPERATIONS MANUAL ba76140e02 10 2017 4320 W CONDUCTIVITY CELL FOR ULTRAPURE WATER ...
Page 2: ...2 ba76140e02 10 2017 4320 Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email environmental ysi com Internet www ysi com Copyright 2017 Xylem Inc ...
Page 3: ...the sensor is connected and is used for measurement and for measured value documentation Storing the calibration data in the sensor ensures that the correct cell constant is automatically used if the sensor is operated with several meters The digital transmission technique guarantees the failure free communi cation with the meter even with long connection cables If the sensor firmware is enhanced ...
Page 4: ...ia the sensor cable to a free IDS connec tor on the meter 4320W via a connecting cable accessory to a free IDS connector on the meter or wireless via an IDS WA S adapter acces sory to a WA capable meter Accessories for the connection of the 4320W sensor to the meter See chapter 6 WEAR PARTS AND ACCESSORIES Opening and closing the IDS plug connection see section 2 2 OPENING AND CLOSING THE IDS PLUG...
Page 5: ...ice step 1 Use your thumb and index finger to press the clips of the connector togeth er and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plug connection is completely dry and clean Align the guidance of the connector with the groove in the plug and insert the connector in the unlocked plug until it catches step 3 Close the locking device step 4 2 4 3 1 ...
Page 6: ...cell Normally the conductivity measuring cell does not age Special measuring media e g strong acids and bases organic solvents or temperatures that are too high may considerably reduce its lifetime or lead to damage The war ranty does not cover failure caused by measuring conditions and mechanical damage Disposal We recommend to dispose of the measuring cell as electronic waste Contamination Clean...
Page 7: ...ment delivers implausible conductivity values Measuring range exceeded Contamination in the area of the electrodes Electrodes damaged Make sure the correct sensor is being used for the application Clean the conductivity measuring cell see section 3 Return the sensor Incorrect temperature display The temperature sensor is not immersed deep enough in the measuring solution Temperature sensor defecti...
Page 8: ...00 cm 1 2 Temperature sensor Integrated NTC 30 30 k at 25 C 77 F Dimensions in mm Weight 4320 Approx 90 g without cable 4320W Approx 95 g Materials Shaft Stainless steel 1 4571 Connection head 4320 POM 4320W POM Conductivity electrodes Stainless steel 1 4571 Thermistor enclosure Stainless steel 1 4571 12 0 12 0 120 0 120 0 176 0 222 8 21 7 21 7 4320 4320W ...
Page 9: ...tance Sensor with connection cable IP 68 2 x 105 Pa or 2 bar Cable plug IP 67 when plugged in The 4320 W meets the requirements according to article 3 3 of the direc tive 97 23 EC pressure equipment directive Measurement con ditions Conductivity measuring range 0 01 μS cm 200 μS cm Temperature range 5 70 C 100 C 23 158 F 212 F Max admissible overpres sure 2 x 105 Pa 2 bar Minimum depth of immer si...
Page 10: ...of the final value display after 20 s Accuracy of the temperature sensor 0 2 K Measuring ranges resolution Measured parameter Measuring range Resolution ӕ µS cm 0 01 19 99 0 0 199 9 0 01 0 1 ρ resistivity kOhm cm 5 00 19 99 20 0 199 9 200 1999 0 01 0 1 1 ρ resistivity MOhm cm 2 00 19 99 20 0 199 9 0 01 0 1 T C 5 0 100 0 0 1 Measured parameter Accuracy 1 digit ӕ ρ 0 5 of measured value T C 0 1 ...
Page 11: ...4320W meter Description Model Order no IDS connection cable 1 5 m IDS CABLE 1 5 903 850Y IDS connection cable 3 m IDS CABLE 3 903 851Y Radio connection 4320W meter Description Model Order no WA capable IDS meter radio module for IDS meter see Internet Radio module for plug head sensor IDS WA S 108 141Y ...
Page 12: ...Wear parts and accessories 4320 W 12 ba76140e02 10 2017 ...
Page 13: ...product for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Product Return form may be downloaded at www ysi com and clicking on the Support tab Telephone 800 897 4151 937 767 7241 Monday through Friday 8 00 AM to 5 00 PM ET Fax 937 767 1058 Email environmental ysi com Mail...
Page 14: ...Contact Information 4320 14 ba76140e02 10 2017 ...
Page 15: ......
Page 16: ...and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise backed by a legacy of innovation For more information on how Xylem can help you go to www xyleminc ...