YSI MultiLab IDS 4120 Operation Manual Download Page 1

OPERATIONS MANUAL

ba76136e04 10/2017

4120(W) / 4130(W)

PH COMBINATION ELECTRODES WITH REFILLABLE LIQUID REFERENCE SYSTEM

Summary of Contents for MultiLab IDS 4120

Page 1: ...OPERATIONS MANUAL ba76136e04 10 2017 4120 W 4130 W PH COMBINATION ELECTRODES WITH REFILLABLE LIQUID REFERENCE SYSTEM ...

Page 2: ...2 ba76136e04 10 2017 4120 W 4130 W Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email environmental ysi com Internet www ysi com Copyright 2017 Xylem Inc ...

Page 3: ... be used with one meter without the need to recalibrate The digital transmission technique guarantees the failure free communica tion with the meter even with long connection cables The sensor firmware can be updated via the meter 2 Commissioning measuring calibration 2 1 Commissioning Scope of delivery Electrode 4120 W 4130 W Operating manual Commissioning The electrode is filled with reference e...

Page 4: ...o IDS plug variants 41x0W Please keep the watering cap It is required for the electrode to be stored Always keep the watering cap clean 4120 4130 via the sensor cable to a free IDS connec tor on the meter 4120W 4130W via a connecting cable accessory to a free IDS connector on the meter or wireless via an IDS WA S adapter acces sory to a WA capable meter Accessories for the connection of the 41x0W ...

Page 5: ...the plug connection Make sure that the plug connection is completely dry and clean Align the guidance of the connector with the groove in the plug and insert the connector in the unlocked plug until it catches step 3 Close the locking device step 4 2 3 Calibration and measurement General rules Make sure the refilling opening for the reference electrolyte solution is open Avoid the carryover of any...

Page 6: ... the same stirring conditions for measuring as for calibrating Subsequent calibrations The frequency of subsequent calibrations depends on the application The meters provide an option where you can enter a calibration interval After the calibration interval has expired the meter will automatically remind you of the due calibration 3 Storage During short measuring breaks With the refilling opening ...

Page 7: ...s and mechanical damage 5 Maintenance and cleaning Refilling the reference electrolyte During operation a small amount of reference electrolyte leaks through the junction from the electrode into the test sample If the level of reference elec trolyte becomes too low with time refill it through the refilling opening Refill ing is very easy with the enclosed dropping bottle Proceed as follows Cut off...

Page 8: ... it in reference elec trolyte solution for at least 1 hour Then recalibrate the electrode 6 Technical data Contamination Cleaning procedure Fat and oil Rinse with water containing household wash ing up liquid Lime and hydroxide deposits Rinse with citric acid 10 by weight Measurement pH measuring range 0 000 14 000 Allowed temperature range C 0 80 C 4120 W 0 100 C 4130 W Allowed temperature range ...

Page 9: ...t 4 pins Shaft dimensions material Shaft length 120 mm 4120 W Shaft diameter 12 mm 4120 W Shaft material Polyamide 4120 W Glass 4130 W IDS plug Synthetic materials Glass fiber reinforced Noryl TPU TPC ET POM PVC PEEK PBT O ring FPM Contacts gold Plated IDS plug Connection type 4 Pole watertight plug connection with lock reverse polarity protected Storage With watering cap filled with KCl 3 mol L A...

Page 10: ...tact your supplier Maintenance equipment Description Model Order no Reference electrolyte solution 250 ml to fill the watering cap KCl 3 mol l Ag free KCl 250 109 705Y Connection cable 4120 W 4130 W meter Description Model Order no IDS connection cable 1 5 m IDS CABLE 1 5 903 850Y IDS connection cable 3 m IDS CABLE 3 903 851Y Radio connection 4120 W 4130 W meter Description Model Order no WA capab...

Page 11: ...product for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Product Return form may be downloaded at www ysi com and clicking on the Support tab Telephone 800 897 4151 937 767 7241 Monday through Friday 8 00 AM to 5 00 PM ET Fax 937 767 1058 Email environmental ysi com Mail...

Page 12: ...Contact Information 4120 W 4130 W 12 ba76136e04 10 2017 ...

Page 13: ......

Page 14: ...and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise backed by a legacy of innovation For more information on how Xylem can help you go to www xyleminc ...