YSI MultiLab 4010P-1W Operating Manual Download Page 18

Operation

MultiLab 4010-1W

18

ba76141e07

08/2018

4.1.4

Socket field

4.1.5

Sensor info

You can display the current sensor data and sensor settings of a connected 
sensor at any time. The sensor data are available in the measured value 
display with the 

<F1>

/[

Info

] softkey.

1 IDS sensors: (pH, ORP, conductivity, D.O.)

2 USB B (device) interface

3 Stirrer (interface for BOD sensor)

4 Power pack

5 Service interface

CAUTION

Only connect sensors to the meter that cannot return any volt-
ages or currents that are not allowed (> SELV and > current 
circuit with current limiting). 
YSI IDS sensors and IDS adapters meet these requirements.

1.

In the measured value display:
Display the sensor data (sensor name, series number) with [

<F1>

Info

].

1

3

5

4

2

25.0

7.007

pH

°C

0

14

01.02.2014 

08:00

i

USB output

Summary of Contents for MultiLab 4010P-1W

Page 1: ...OPERATING MANUAL ba76141e07 08 2018 MultiLab 4010 P 1 W DIGITAL METER FOR WIRELESS IDS SENSORS 4010P 1W 4010 1W 4010 1W 4010P 1W IDS WA IDS WA...

Page 2: ...4010 1W Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email info ysi com Internet www ysi com Copyright 2018 Xylem Inc For the most recent version of the ma...

Page 3: ...operation 12 2 2 1 Authorized use 12 2 2 2 Requirements for safe operation 12 2 2 3 Unauthorized use 12 3 Commissioning 13 3 1 Scope of delivery 13 3 2 Power supply 13 3 3 Initial commissioning 13 3...

Page 4: ...ration 43 7 Dissolved oxygen 44 7 1 Measuring 44 7 1 1 Measuring D O 44 7 1 2 Measuring the temperature 46 7 2 FDO Check checking the FDO 4410 47 7 2 1 Why should you check the sensor 47 7 2 2 When sh...

Page 5: ...nt settings 73 9 5 1 System 73 9 5 2 Data storage 74 9 5 3 Automatic Stability control 74 9 5 4 Automatic switch off function 74 9 5 5 Display illumination 75 9 6 Reset 75 9 6 1 Resetting the measurem...

Page 6: ...4 3 Conductivity 94 14 4 General topics 95 15 Technical data 96 15 1 Measuring ranges resolution accuracy 96 15 2 General data 96 16 Firmware update 98 16 1 Firmware update for the meter MultiLab 4010...

Page 7: ...asurements quickly and reliably The MultiLab 4010 1W provides the maximum degree of operating comfort reliability and measuring certainty for all applications The MultiLab 4010 1W supports you in your...

Page 8: ...sensors 1 3 1 IDS sensors IDS sensors support the automatic sensor recognition show only the settings relevant to the specific sensor in the setting menu process signals in the sensor digitally so tha...

Page 9: ...eplace the sensor cable with an energy saving Bluetooth LE radio connection 1 3 3 Automatic sensor recognition The automatic sensor recognition for IDS sensors allows to operate an IDS sensor with dif...

Page 10: ...atic sensor recognition a meter that supports the automatic sensor recognition e g MultiLab 4010 1W and a digital IDS sensor are required In digital IDS sensors sensor data are stored that clearly ide...

Page 11: ...ods might be damaged if the actions mentioned are not taken 2 1 2 Safety signs on the meter Note all labels information signs and safety symbols on the meter and in the battery compartment A warning s...

Page 12: ...eration Note the following points for safe operation The meter may only be operated according to the authorized use specified above The meter may only be supplied with power by the energy sources ment...

Page 13: ...Lab 4010 1W is supplied with power in the following ways Line power operation via the supplied power pack A sensor with stirrer can only be operated with line power Battery operation 4 batteries 1 5 V...

Page 14: ...d 1 Open the battery compartment 1 on the underside of the meter 1 Battery compartment CAUTION Make sure that the poles of the batteries are positioned cor rectly The signs on the batteries must corre...

Page 15: ...ine voltage at the operating site must lie within the input voltage range of the original power pack see section 15 2 GENERAL DATA page 96 CAUTION Use original power packs only see section 15 2 GENERA...

Page 16: ...nformation on a sensor On Off Switches the meter on or off M Selects the measured parameter Quits the settings CAL CAL__ Calls up the calibration procedure Displays the calibration data STO STO_ Saves...

Page 17: ...Sensor symbol calibration evaluation calibration interval 6 Measured temperature with unit 7 Status information meter 8 Softkeys and date time AR Stability control AutoRead is active HOLD Measured val...

Page 18: ...ORP conductivity D O 2 USB B device interface 3 Stirrer interface for BOD sensor 4 Power pack 5 Service interface CAUTION Only connect sensors to the meter that cannot return any volt ages or currents...

Page 19: ...he meter performs a self test 2 Connect the sensor The meter is ready to measure 1 Switch the meter off with On Off 25 0 7 007 pH C 0 14 More 4110 B09250001 pH resolution 0 001 mV resolution 0 1 Buffe...

Page 20: ...ts The selection is done with the keys The current selection is dis played with a frame Submenus The name of the submenu is displayed at the upper edge of the frame Sub menus are opened by confirming...

Page 21: ...ing with ENTER Example Display the Calibration record function Messages Information is marked by the i symbol It cannot be selected Example General Interface Clock function Service information Reset S...

Page 22: ...he instrument is in the setting mode 3 Select the System submenu with The current selection is displayed with a frame 4 Open the System submenu with ENTER Calibration record Calibration data storage B...

Page 23: ...ting with ENTER The meter switches to the measuring mode The selected language is active General Interface Clock function Service information Reset System 03 04 2013 08 00 Back Language Deutsch Audio...

Page 24: ...e time and date format The date format can be switched from the display of day month year dd mm yy to the display of month day year mm dd yy or mm dd yy The date and time are reset to default after a...

Page 25: ...ith and ENTER The time is set 7 If necessary set the Date and Date format The setting is made similarly to that of the time 8 To make further settings switch to the next higher menu level with F1 or S...

Page 26: ...y control is switched off You can start the Stability control manually at any time irrespective of the set ting for automatic Stability control see section 9 5 3 AUTOMATIC STABILITY CON TROL page 74 i...

Page 27: ...and the display of the mea sured parameter flashes The HOLD AR status indicator appears as soon as a stable measured value is recognized The progress bar disappears the display of the mea sured parame...

Page 28: ...eter see section 9 1 2 BUFFER SETS FOR CALIBRATION page 64 Manual calibration ConCal Any buffer solutions can be used The nominal values of the buffer solutions are manually entered by the user during...

Page 29: ...l The AR status indicator is displayed The measured parameter flashes 7 Wait for the end of the measurement with stability control beep HOLD AR status indicator or take over the calibrated value with...

Page 30: ...l beep HOLD AR status indicator or take over the calibrated value with ENTER The calibration display for the next buffer appears voltage display 13 If necessary finish the calibration procedure as a t...

Page 31: ...set has been measured Then the calibration record is dis played The calibration line is determined by linear regression If single point calibration was set in the menu the calibration proce dure is au...

Page 32: ...display 10 If necessary finish the calibration procedure as a single point calibration with M The calibration record is displayed For single point calibration the instrument uses the Nernst slope 59...

Page 33: ...necessary finish the calibration procedure as a two point calibration with M The calibration record is displayed pH 4 035 24 8 C ConCal 2 01 07 2011 08 00 Buffer 18 Thoroughly rinse the sensor with de...

Page 34: ...displayed or Continue calibrating using the next buffer with ENTER After the fifth buffer has been measured the calibration is automat ically finished Then the calibration record is displayed The cali...

Page 35: ...en into account The evaluation appears on the display and in the calibration record Menu item Setting function Explanation Calibration Calibration data stor age Display Displays the calibration record...

Page 36: ...S sensors you can optionally enable a more finely graded calibration evaluation QSC see section 5 2 9 QSC FUNCTION SENSOR QUALITY CONTROL page 38 Display Calibration record Zero point mV Slope mV pH 4...

Page 37: ...value needle 2 Marking lines for all nominal buffer values used with the last valid cali bration 3 Measuring range for which a valid calibration is available Measured values in this range are suitable...

Page 38: ...n with special QSC buffer solutions It covers the measuring range of the sensor pH 2 to pH 11 The QSC initial calibration determines the actual condi tion of the sensor and stores it as a reference in...

Page 39: ...al calibration very carefully It determines the reference value for the sensor This reference value cannot be overwritten or reset As soon as the QSC function was enabled it is not possible to return...

Page 40: ...the buffer solutions of the QSC Kit 2 Open the menu for measurement settings with ENTER 3 In the QSC menu select Control calibration with The calibration display appears AutoCal QSC Kit is displayed...

Page 41: ...n as the measured value is outside the stability range when the automatic Stability control is switched off You can start the Stability control manually at any time irrespective of the set ting for au...

Page 42: ...ed and the display of the mea sured parameter flashes The HOLD AR status indicator appears as soon as a stable measured value is recognized The progress bar disappears the display of the mea sured par...

Page 43: ...ed temperature sensor is operated the tempera ture value is displayed and recorded together with the measured value 6 2 ORP calibration ORP electrodes are not calibrated You can however check ORP elec...

Page 44: ...FDO Check instead 7 1 Measuring 7 1 1 Measuring D O Selecting the displayed measured parameter You can switch between the following displays with M D O concentration mg l D O saturation A sensor with...

Page 45: ...asurement signal The stability has a considerable impact on the reproduc ibility of measured values The measured parameter flashes on the display as soon as the measured value is outside the stability...

Page 46: ...tus indicator appears as soon as a stable measured value is recognized The progress bar disappears the display of the mea sured parameter stops flashing and there is a beep The current measurement dat...

Page 47: ...sensor cap is contaminated or at the end of its lifetime After the sensor cap was exchanged Routinely within the framework of the company quality assurance 7 2 3 Perform a FDO Check Proceed as follow...

Page 48: ...e D O saturation abbreviated saturation is 100 Therefore the validity scope is 98 102 The check resulted in a measured value of 99 3 The measurement error is within the specified validity scope No cle...

Page 49: ...ISO 5813 At the same time the relative slope is adapted to the comparison measurement by a correction multiplier When the correction multiplier is active the Factor indicator appears in the mea surin...

Page 50: ...O sensor into the calibration vessel 3 For FDO 4410 with the check and storage vessel FDO Check Close the check and storage vessel FDO Check Leave the sensor in the calibration vessel or check and st...

Page 51: ...he AR status indicator is displayed The measured parameter flashes 5 Wait for the measurement with stability control to be finished beep HOLD AR status indicator 6 Using set the measured value to adju...

Page 52: ...for this sensor 1 Connect the D O sensor to the meter 2 Place the D O sensor in a solution that does not contain any dissolved oxygen A solution not containing any dissolved oxygen can be prepared by...

Page 53: ...e The zero point is calibrated The ZeroCal status indicator is displayed 7 Carry out a calibration see section 7 3 3 CALIBRATION PROCEDURES page 49 The calibration record is automatically transmitted...

Page 54: ...as determined is compared to the characteristic curve of an ideal sen sor at similar environmental conditions relative slope S S Ssen sor Sideal sensor An ideal sensor has a slope of 1 Display Calibra...

Page 55: ...st this multiplier to meet your requirements in the range 0 40 1 00 The multiplier is set in the menu for the parameter TDS Stability control AutoRead HOLD function The stability control function Auto...

Page 56: ...Using ENTER activate the Stability control function manually The AR status indicator appears while the measured value is assessed as not stable A progress bar is displayed and the display of the measu...

Page 57: ...n methods Nonlinear temperature compensation nLF according to EN 27 888 Linear temperature compensation Lin with adjustable coefficient No temperature compensation off Application tips Select the foll...

Page 58: ...ocedure for IDS conductivity sensors with a cell constant in the range 0 450 0 500 cm 1 Determining the cell constant Calibration with freely selectable check and calibration standard Comprehensive an...

Page 59: ...t of the IDS sensor e g 4310 0 475 cm 1 For this calibration procedure the Type setting must be set to cal Proceed as follows to determine the cell constant 1 Connect the conductivity sensor to the me...

Page 60: ...the cell constant 1 Connect the conductivity sensor to the meter 2 In the measured value display select the conductivity parameter with M 3 Open the menu for measurement settings with ENTER 4 Select...

Page 61: ...alibration record is automatically transmitted to the interface after calibrating Menu item Setting function Explanation Calibration Calibration data storage Display Displays the calibration records F...

Page 62: ...Conductivity MultiLab 4010 1W 62 ba76141e07 08 2018 Calibration record USB output CALIBRATION Cond 03 04 2013 07 43 33 4310 Ser no 09250033 Cell constant 0 476 1 cm 25 0 C Sensor...

Page 63: ...ee section 9 1 2 BUFFER SETS FOR CALIBRATION page 64 and section 5 2 PH CALIBRATION page 28 Calibration Single point calibration yes no Quick calibration with 1 buffer Calibration Calibration interval...

Page 64: ...tion 0 1 1 Resolution of the mV display Reset Resets all sensor settings to the delivery condition see sec tion 9 6 1 RESETTING THE MEASUREMENT SETTINGS page 75 Menu item Possible setting Explanation...

Page 65: ...995 4 005 7 002 9 208 25 C 13 Fisher 2 007 4 002 7 004 10 002 25 C 14 Fluka BS 4 006 6 984 8 957 25 C 15 Radiometer 1 678 4 005 7 000 9 180 25 C 16 Baker 4 006 6 991 10 008 25 C 17 Metrohm 3 996 7 00...

Page 66: ...the calibra tion interval The calibration interval is set to 7 days in the factory You can change the interval 1 999 days 22 Reagecon 20 2 000 4 000 7 000 10 000 13 000 20 C 23 Reagecon 25 2 000 4 00...

Page 67: ...sured value display and press the ENTER key After completing the settings switch to the measured value display with M The possible settings are individually displayed for each sensor Below the set tin...

Page 68: ...ter exchanging the sensor cap For details see section 9 3 2 ENTER SENSOR CAP COEFFICIENTS PROOBOD page 69 Calibration Membrane cap only for 4100 Pro BOD 5010 with 4011 Adapter Black Yellow After excha...

Page 69: ...r FDO 4410 30 300 s Response time of the signal filter in seconds A signal filter in the sensor reduces the limits of variation of the mea sured value The signal filter is characterized by the respons...

Page 70: ...t settings are printed in bold General setting menu for conductivity Menu item Possible setting Explanation Calibration Calibration record Displays the calibration record of the last calibration Calib...

Page 71: ...adjustable cell con stant This menu item is only available when Type man is set The current conductivity value is displayed in the status line Temp comp TC Method nLF Lin off Procedure for temperatur...

Page 72: ...ion see section 8 2 TEM PERATURE COMPENSATION page 57 This setting is only available for the measured parameters conductivity and resistivity Temp comp TC Linear coeff 0 000 2 000 3 000 K Coefficient...

Page 73: ...me 10 min 1h 24 h Adjusts the switch off time System General Tem perature unit C F Temperature unit degrees Celsius or degrees Fahrenheit All temperature values are displayed with the selected unit Sy...

Page 74: ...switch off function in order to save the batteries see section 9 5 1 SYSTEM page 73 The automatic switchoff function switches off the meter if no key is pressed for an adjustable period The automatic...

Page 75: ...pH The following settings for pH measurements are reset to the default settings with the Reset function The sensor settings are reset under the Reset menu item in the menu for cali bration and measure...

Page 76: ...pen the settings display the required measured parameter in the measured value display and press the ENTER key Conductivity The following settings for conductivity measurements are reset to the defaul...

Page 77: ...6 2 Resetting the system settings The following system settings can be reset to the default condition The resetting of the system settings is done in the menu Storage config System Reset To open the...

Page 78: ...he same time output to the interface When the storage is full When all storage locations are occupied it is not possible to continue storing Then you can e g transmit the stored data to a PC see secti...

Page 79: ...al display of the storage usage ID number 1 Interval 30 s Duration 180 min continue 0d03h00min 0 1d17h33min Automatic data storage 1 2 3 03 04 2013 08 00 Back Menu item Possible setting Explanation ID...

Page 80: ...as follows to switch off the automatic data storage function before the adjusted storage duration has expired 1 Remaining storing duration 2 Graphical display of the storage duration If the automatic...

Page 81: ...e settings are explained here using the manual data storage as an example The same settings and functions are available for the automatic data storage Menu item Setting function Explanation Data stora...

Page 82: ...a storage see section 10 3 1 MANAGING THE MEASUREMENT DATA STORAGE page 81 10 3 3 Measurement dataset A complete dataset consists of Date time Meter name series number Sensor name series number ID num...

Page 83: ...storing stable measured value Otherwise the AR dis play is missing Calibration evaluation 4 levels or no evaluation or QSC percentage 10 3 4 Storage locations The MultiLab 4010 1W meter has two measur...

Page 84: ...he measurement data on the PC e g MultiLab Importer see CD ROM or Internet Installation of the USB driver Connecting a PC Adjusting the set tings for the data transmission 1 Insert the supplied instal...

Page 85: ...l Simultaneously with every automatic data storing process see section 10 2 AUTOMATIC DATA STORAGE AT INTER VALS page 79 Stored mea sured values Manual Displayed dataset with F2 USB out put after call...

Page 86: ...08 2018 11 3 BOD Analyst Pro With the aid of the BOD Analyst Pro software you can administrate BOD measurement on the PC and automatically calculate the measured values More detailed information can...

Page 87: ...d ON OFF 3 Lever to open the paper compartment 4 Lid of the paper compartment 2 1 3 4 1 Connect the power pack to the MultiLab 4010P 1W The LED lights up green The printer is ready for operation or Wh...

Page 88: ...Changing the roll of paper thermal paper Menu item Possible set ting Explanation System Printer Font size 12x20 8x16 7x16 Select the font size for the printer A print sample font of the printer to vi...

Page 89: ...able from the meter Function store automatically at intervals with long intervals is switched on Switch off the function see section 10 2 AUTOMATIC DATA STORAGE AT INTERVALS page 79 No paper available...

Page 90: ...on maintenance You can operate the meter either with normal batteries or with rechargeable batteries Ni MH In order to charge the batteries an external charging device is required 1 Open the battery c...

Page 91: ...tem statutory in your country If you have any questions please contact your supplier 4 Set the date and time see section 4 4 5 EXAMPLE 2 ON NAVIGATION SETTING THE DATE AND TIME page 24 Dispose of used...

Page 92: ...Air bubble in front of the junction Remove air bubble e g sway or stir the solution Cable broken Replace the sensor Gel electrolyte dried out Replace the sensor Cause Remedy IDS pH sensor The values d...

Page 93: ...H sensor test sample Conductivity too low Use suitable IDS pH sensor Temperature too high Use suitable IDS pH sensor Organic liquids Use suitable IDS pH sensor Cause Remedy IDS pH sensor Sensor unsuit...

Page 94: ...sensor cap Replace the sensor Calibration failed Recalibrate D O concentration during zero point calibration too high Immerse the sensor in an oxygen free solution More information and instructions o...

Page 95: ...eries almost empty Replace the batteries see section 13 1 MAINTENANCE page 90 Cause Remedy Operating condition undefined or EMC load unallowed Processor reset Press the ENTER and On Off key simultaneo...

Page 96: ...cal safety Protective class III Test certificates CE Ambient conditions Storage 25 C 65 C Operation 5 C 55 C With the power pack connected 5 C 40 C Admissible relative humidity Yearly mean 75 30 days...

Page 97: ...te Adjustable 1200 2400 4800 9600 19200 Baud Data bits 8 Stop bits 2 Parity None Handshake RTS CTS Cable length max 3 m 9 843 feet Guidelines and norms used EMC EC directive 2014 30 EC EN 61326 1 EN 6...

Page 98: ...VERSION IS IN THE METER OR IDS SENSOR PAGE 95 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder already exists for the meter o...

Page 99: ...METER OR IDS SENSOR PAGE 95 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder already exists for the sensor or sensor type the...

Page 100: ...ty divided by the unit of the molality The practical pH value is the value of a pH measurement Potentiometry Name of a measuring technique The signal depending on the mea sured parameter of the electr...

Page 101: ...TRef D O saturation Short name for the relative D O saturation Relation of the D O partial pressure in the test sample to the D O partial pressure of air at the current air pressure Example 100 means...

Page 102: ...l settings of a measuring system Resolution Smallest difference between two measured values that can be dis played by a meter Stability control AutoRead Function to control the measured value stabilit...

Page 103: ...Firmware update 98 I Initial commissioning 13 Initialize 75 Interval for calibration 66 K Keys 16 M Measured value display 20 Measurement accuracy 66 Measurement data storage Storage locations 83 Mea...

Page 104: ...Index MultiLab 4010 1W 104 ba76141e07 08 2018 Transmitting data 84 automatic 85 Manual 85 Transmitting measured values 84 Two point calibration pH 30 32 Z Zero point of pH electrode 28...

Page 105: ...MultiLab 4010 1W Index ba76141e07 08 2018 105...

Page 106: ...9 10 531 9 930 9 362 8 826 9 0 11 559 10 907 10 290 9 707 9 156 8 636 10 0 11 288 10 656 10 058 9 493 8 959 8 454 11 0 10 027 10 415 9 835 9 287 8 769 8 279 12 0 10 777 10 183 9 621 9 089 8 586 8 111...

Page 107: ...5 842 5 576 5 321 5 078 41 0 6 312 6 026 5 753 5 493 5 243 5 005 42 0 6 213 5 934 5 667 5 411 5 167 4 993 43 0 6 116 5 843 5 581 5 331 5 091 4 861 44 0 6 021 5 753 5 497 5 252 5 017 4 793 45 0 5 927 5...

Page 108: ...33 668 8 89 2 892 3496 1066 88 26 03 661 2 88 1 881 3804 1106 87 25 73 653 6 87 2 871 4115 1254 86 25 43 646 86 1 861 4430 1350 85 25 13 638 4 85 1 851 4747 1447 84 24 84 630 8 84 1 841 5067 1544 83 2...

Page 109: ...he multiplier for a specific sample 1 Determine the specific conductance of a water sample from the site 2 Filter a portion of water from the site 3 Carefully measure a volume of the filtered water Co...

Page 110: ...Appendix MultiLab 4010 1W 110 ba76141e07 08 2018...

Page 111: ...returning a product for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Pr...

Page 112: ...Contact Information MultiLab 4010 1W 112 ba76141e07 08 2018...

Page 113: ......

Page 114: ...yze and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with...

Reviews: