MultiLab 4010-1W
Dissolved oxygen
ba76141e07
08/2018
53
7.3.7
Calibration data
The calibration record of the last calibration is available in the menu
Calibration
/
Calibration record
. To open it in the measured value display, press the
<CAL__>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in the
measured value display, press the
<ENTER>
key.
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the cal-
6.
Switch to the measured value display with
<F1>
/
[
continue
]
.
The zero point is calibrated.
The [ZeroCal] status indicator is displayed.
7.
Carry out a calibration (see section 7.3.3 C
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/
function
Explanation
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to
the interface with
<F2__>
[USB output]
.
Quit the display with
<F1>
/
[Back] or
<ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data stor-
age to the interface (see section
12 T