![YSI MultiDO 5400 User Manual Download Page 74](http://html.mh-extra.com/html/ysi/multido-5400/multido-5400_user-manual_915316074.webp)
YSI 5X00
147
YSI 5X00
146
5. Scroll and select Relay or 4/20mA output type.
6. 4/20mA outputs test an IOEM 4/20mA output channel. For 4/20mA out-
put test, configure a Milliamp value. The Milliamp raange is 0.00 to 21.00.
7. Select Yes at Energized menu to toggle state of output.
Configuring the 5X00
8. Scroll and select Send Request.
at 5X00 Run Screen to send a com-
mand to the local, or networked, instrument to trigger configured output
test condition.
9. Verify locally, at the instrument whose output is being tested, that output
is at correct state or value.
10.
after completing instrument channel output testing.
Service Hold
Menus → System → Service Hold
Pressing
at the Run Screen puts the 5X00 into Service Hold. The Service
Hold time is user selectable (1-255 sec, min, or hours). Service Hold turns off any
active relays and disables any sensor, aux, and timer system relays from energizing
regardless of value or condition. 5X00 alarm emails are not sent during service hold.
After the service hold time expires, or is cancelled, the sensor systems are reset after
30 seconds. Timer systems are not reset.
1.
Use
▲
and
▼
to scroll and highlight submenus.
2.
Press
to select.
3. Enable and configure submenus.
4. Press
to save configuration and return to Run Screen.
5. Press
at Run Screen to enter Service Hold . See Svr Hold - page
196 for additional information.
Note:
-
Place the 5X00 in service hold as instructed in step 3 of calibration checks in
Maintenance - page 209.
Configuring the 5X00