Step 9 Wire and Install Communication Equipment
Communication equipment must be installed in order to communicate, either locally
or remotely, with the 5X00. The communication connection can be made locally
(RS232) or remotely with a serial to ethernet device.
Only 1 connection method
can be used at a time.
AquaManager supports both connection methods.
Notes:
-
Firmware can only be uploaded to an instrument using a local RS232 connec-
tion. See
Downloader
- page 126.
-
On a network, install and configure RS232 or the serial to ethernet device
hardware on the master node only.
Email Alarming
Email alarming can be configured in 1 of 2 ways:
•
from PC running AutoPoll Program that is included with AquaManager
software (PC must have internet connection). This is referred to as
Aqua-
Manager email
.
•
via a serial to ethernet* device with TCP/IP connection at the master 5X00.
This is referred to as
5X00 email
.
*Serial to ethernet connection requires access to an SMTP server accessible by the
local network.
RS232 Communication
Use Category 5e patch cable from RS232 port at location
J
- figure 3.31 of I/O
Board to PC using RJ45 to DB9 adapter. (Common lettering references are used
for both 5400 and 5500 models).
Directions to Wire Direct RS232 PC Connection
WARNING: Disconnect external power to the unit before wiring.
AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer
un câblage quelconque.
1. Perform steps 1-4 of 5X00 installation and steps 5 - 8 as required for
the application.
2. Open front panel - page 41.
Installation and Wiring
3. Feed direct connect cable through drilled hole of rubber grommet and
to location
J.
CAUTION
: Run high and low voltage cables through separate bulkhead and
conduit.
4. Connect one end of communication cable to RJ45 at location
J
. Connect
other end to RS232 PC com port using RJ45 to DB9 adapter figure 3.31.
B
C
D
F
G
J
L
M
N
O
D
o not c
onnec
t or disc
onnec
t IDC r
ibbon cable
when 5400 is po
w
er
ed
. S
er
ious damage can oc
cur
.
figure 3.31
5. Complete other wiring to configure 5X00 system.
6. Close front panel - page 41.
7. Test RS232 communication.
Notes:
-
When using a USB-Serial Adapter, verify that the port number assigned dur-
ing driver installation is the same port configured when using AquaManager
and/or the Firmware Updater
-
Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS YSI
p/n
773
(64-bit).
-
When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232
Line Extender/Booster. Mount the booster in an area where it is protected
from moisture.
Installation and Wiring