condition. Timer systems are not reset. To reset timer systems, turn power off and
then on at the 5200A. Press
to exit Service Hold.
The Run Screen displays a clock count down of the Service Hold remaining time.
Notes:
-
If a timer is active (relay energized) at service hold, the relay will not reenergize
after service hold expires regardless if timer time has not expired. There is
only one “on” time for all timer systems.
-
Entering Downloader mode places 5200A into Service Hold.
-
“Inst. Service Hold” is displayed on any AquaManager map icon associated
with an instrument in Service Hold. AquaManager alarm emails are not sent
for any alarm condition for a 5200A that is in Service Hold.
-
5200A alarm emails are not sent during service hold.