Service Hold
Menus → System → Service Hold
Pressing
at the Run Screen puts the 5200A into Service Hold. The Service
Hold time is user selectable (1-255 sec, min, or hours). Service Hold turns off any
active relays and disables any sensor, aux, and timer system relays from energizing
regardless of value or condition. 5200A alarm emails are not sent during service
hold. After the service hold time expires or is cancelled, sensor systems are reset
after 30 seconds. Timer systems are not reset.
1. Use
▲
and
▼
to scroll and highlight submenus.
2. Press
to select.
3. Enable and configure submenus.
4. Press
to save configuration and return to Run Screen.
5. Press
at Run Screen to enter Service Hold. See Service Hold But-
ton - page 195 for additional information.
Note:
-
Place the 5200A in service hold as instructed in step 3 of calibration checks
in Maintenance, page 209.
Version
Menus → System →Version
The Version menu contains information about the 5200A including: display and
control board firmware code version, firmware compile date and time, control board
temperature, AC and DC voltage and model number.
Notes:
-
Firmware version information is required when contacting YSI support.
-
The clock chip temperature range starts at 0oC. Values < 0oC are not valid
numbers.
Configuring the 5200A
-
Firmware date and time compilation are displayed below the version number.
-
The normal operating temperature range will be a few degrees above ambient.
-
AC powered units will run at a higher temperature than DC powered units.
-
To view current AC and DC power values press
to return to the Run
Screen and then return to the menu. Power information does not refresh
when the Version menu is open.
AquaManager and the System Menu
AquaManager users can configure most 5200A system menus from a PC. After poll-
ing the 5200A, use “Systems Management” from the node window to view and upload
5200A configurations. System menus are configured at the Setup and Input tabs.
Configuring the 5200A