Test RS232 Communications
Apply power to 5200A. Insert node into AquaManager program. Poll node. If
wired correctly, AquaManager will get 5200A configuration and logs.
Ethernet (TCP/IP) Module Communication
The optional serial to ethernet device must be installed on the 5200A control board
when using TCP/IP for communication. Complete installation instructions are
provided with the optional serial to ethernet device.
Notes:
-
See Menus → System → Communications → Ethernet beginning on
page
for configuration information
.
-
An SMTP server accessible by the local network must be provided by your
local IT support for ethernet connection.
Consult the network administrator to verify compatibility of ethernet configuration.
Step 9 Connect Probe/Cable Assembly
Either a YSI 5561(DO & temp) or YSI 5562 (DO, temp, conductivity, pH & ORP)
probe/cable assembly can be used with the 5200A. Both probe/cable assemblies
have an MS19 connector (Military Spec). Connect the probe/cable assembly as
described below.
Directions to Connect Probe/Cable Assembly
1. Perform steps 1 - 4 of 5200A installation and steps 5 - 8 as required for
the application.
2. Line up pins and guides with the holes and indentations on the cable con-
nector at the bottom of the YSI 5200A - figure 3.30.
3. Hold the cable firmly against the cable connector; then, turn the locking
mechanism clockwise until it snaps into place. (Remove the cable from
the instrument by turning the cable connector counterclockwise until the
cable disengages from the instrument.)
Installation and Wiring
Figure 3.30
Notes:
-
Calibrate sensors prior to use. See calibration - page 75.
-
CAUTION: Do not ground the probe body.
Step 10 Install and Wire IOEM/REM Instruments
1. Perform steps 1-4 of 5X00 installation and steps 5 - 8 as required for
the
application.
Note:
-
At step 7, Wire Network - page 52, IOEM and REM instruments are always
slave devices on a network.
2. Follow all installation and wiring steps as required for application as
provided at Chapter 3 of separate manual “IOEM and REM Safety and Instal-
lation Guide.”
Lightning and Surge Protection
AC line voltage surge suppressors protect field equipment on any AC line-to-ground
from damage due to electrical transients induced in the interconnecting power lines
from lightning discharges and other high voltage surges. Surge protection devices
are strongly recommended to protect your equipment from secondary surges and
lightning on outdoor installations. Follow the recommendations provided when
choosing and incorporating surge protection devices into your operation:
Installation and Wiring