![YSI 3100 Operation Manual Download Page 41](http://html.mh-extra.com/html/ysi/3100/3100_operation-manual_915296041.webp)
37
11. Appendix A - Specifications
Modes
Conductivity
Temperature compensated conductivity
Salinity
Temperature
Conductivity
Range
Accuracy
Resolution
Frequency
0 - 49.99
µ
S*
±
0.5% full scale
0.01
µ
S
70 Hz
0 - 499.9
µ
S
±
0.5% full scale
0.1
µ
S
70 Hz
0 - 4999
µ
S**
±
0.5% full scale
1
µ
S
240 Hz
0 - 49.99 mS***
±
0.5% full scale
0.01 mS
1562 Hz
0 - 499.9 mS****
±
0.5% full scale
0.1 mS
1562 Hz
Salinity
Range
Accuracy
Resolution
0-80 ppt (NaCl)
±
2% or
±
0.1 ppt
0.1 ppt
Temperature
Range
Accuracy
Resolution
-5 - 95°C
±
0.1°C +1LSD
0.1°C
Temperature Compensation
Method
Linear
Ref. temp., °C
15 - 25
Temp. Coefficient
0 - 4%/°C
Cell constant, cm
-1
0.01
0.08 - 0.12
0.8 - 1.2
8 - 12
Display
LCD
Cell connector
7-pin Mini Din
Power
AC, 115V, 220V
Approvals
UL, CSA, CE
Environmental requirements
95% RH non-cond
Size
9 x 9.5 x 4.4 inches
22.9 x 24.1 x 11.2 cm
Weight
2.6 pounds
1.1 kg
* Requires a cell constant of K=0.01, K=0.1 or K=1.
** Requires a cell constant of K=0.1, K=1 or K=10.
*** Requires a cell constant of K=1 or K=10.
**** Requires a cell constant of K=10.