7.2.5 Monitor
From the Service menu, touch the [Monitor] tab.
Test Outputs
Touch the DAC A or DAC B [0.00V] button to test the analog output by setting the output to 0, 5 or 10VDC.
Touch the Ready, Feed Pump, Filtrate Pump, or SysErr [Off] button and change it to [On] to test the corresponding output.
Valve
Touch the Opened Valve [Antiseptic] button to change the state of the valve and change it to [Sample]
Prime Monitor Pump
Touch the Pump Sample [Off] button and change it to [On] to prime the monitor sample line. The pump will turn off
automatically.
If you are using an antiseptic solution, touch the Pump Antiseptic [Off] button and change it to [On] to prime the antiseptic
solution. The pump will turn off automatically.
NOTE: After sterilization, the antiseptic solution must be primed immediately after the tubing is reconnected.