7.1.4 Auto-Cal Settings
From the Configuration screen, touch the [Auto] tab to display the Auto-Cal and Auto-Flush settings.
Touch the [Calibration] button for a module to display the Calibration settings for that module.
The default Auto-calibration settings are shown above. You may alter any of these parameters to suit your application,
however,
you may compromise precision and/or accuracy
when doing so. YSI’s stated specifications are based on the
default settings. These selections are provided as part of the overall concept of the 2900 Series flexibility.
To change the value of a Calibration parameter, simply touch the value to open the numeric keypad. Enter the new value,
then touch [OK].
NOTE:
To disable a Calibration parameter, touch the [ON] button and change it to [OFF].
With AutoCal set to [On], the instrument will keep the module calibrated while idle in the Run screen. If you only want the
instrument to calibrate while running samples/plates, change the AutoCal setting to [Off].
Touch [Save] to save your changes.