The Remote Communication Port
Remote Communication Port
Remote Communication Port
Remote Communication Port is a 25 pin RS232C serial port. It is used to
interface with personal computers.
The Power Switch
Power Switch
Power Switch
Power Switch is a rocker switch that allows you to control power to the
instrument.
The Infrared Printer Port
Infrared Printer Port
Infrared Printer Port
Infrared Printer Port is used to transmit data to the optional battery
powered printer.
The Power Receptacle
Power Receptacle
Power Receptacle
Power Receptacle is a DC power inlet. The battery charger supplied with
the SPORT is connected here to recharge the instrument battery.
IMPORTANT: Probe polarization will be disrupted when power is turned off.
Time for restabilization will be needed when power is restored.
3.3 Reagent Preparation
If you are going to use only the manual mode of operation, the only solution you
will need to prepare is the Buffer solution as the Reference solution is only used
as part of the Automatic mode.
You can prepare the reagent solutions for the SPORT yourself. These reagents,
along with a mixing bottle and a graduated syringe (3cc), are provided in the
YSI Model 1504 Starter Kit which must be ordered separately. Please note that
the reagents in the starter kit may also be ordered separately.
IMPORTANT: The mixing bottle and Syringe must be cleaned every week. See
Section 8.2 Preventative Maintenance.
If your sample measuring protocol requires the total lysis of blood cells, or if
you are going to preserve the samples in YSI 2372 Preservative Tubes for later
analysis, you must refer to Appendix D Sample Preservation, for further
information before proceeding.
Automatic Mode
Automatic Mode
Automatic Mode
Automatic Mode
Buffer Solution
Buffer Solution
Buffer Solution
Buffer Solution
Fill the 500mL mixing bottle with reagent water. Reagent water must be very
pure. Use high quality deionized or distilled water. Add one package of YSI
2357 Buffer Concentrate and stir, waiting long enough to be certain that the
buffer chemicals have completely dissolved.
Pour buffer solution into the instrument Buffer Bottle. Put the cap back on the
Buffer bottle and place it back into the instrument.
Summary of Contents for 1500 Sport
Page 2: ......
Page 8: ...1 1 1 1 2 2 2 2 ...
Page 46: ...5 5 5 5 1 1 1 12 2 2 2 ...
Page 68: ...6 6 6 6 22 22 22 22 ...
Page 78: ...7 7 7 7 10 10 10 10 ...
Page 86: ......
Page 100: ...9 9 9 9 14 14 14 14 ...
Page 114: ...12 12 12 12 4 4 4 4 ...
Page 116: ...13 13 13 13 2 2 2 2 ...
Page 122: ...14 14 14 14 6 6 6 6 ...
Page 124: ...15 15 15 15 2 2 2 2 ...
Page 126: ...16 16 16 16 2 2 2 2 ...
Page 128: ...17 17 17 17 2 2 2 2 ...
Page 132: ...18 18 18 18 4 4 4 4 ...