37
Quality control test with mylife
™
ControlGDH control solution
Example of control
solution range
printed on your test
strip vial label.
A quality control test should be performed whenever you want to
check if your BGMS is working properly or whenever you want to
practise and check the correct the testing procedure.
Only use a control solution tested with the BGMS under control
solution mode. If the test result is within the control solution range
printed on the label of the test strip vial, the BGMS passes the
quality control test. That means your BGMS is working correctly.
There are two different types of control solutions available: normal
and high.
Control
solution
range
Normal
High
mmol/L
4.6 – 6.3
13.2 – 17.8
mg/dL
83 – 113
237 – 321