3.5.3 Remote Sensor (Style: S2)
IMPORTANT
The following types can not be changed direction of
electrical connection after delivery.
• Submersible Type (Model: AXW
W)
• For District Heating and Cooling or Condensation-
proof (Optional code: DHC)
(1) The following tools are required.
• Hexagonal wrench (nominal size 5)
(2) Turn off the power to the flowmeter.
(3) Using the wrench, unscrew the four hexagonal bolts
at the neck of remote sensor.
F0327.ai
AXW Remote Sensor
CAUTION
Do not loosen the screw at the opposite side of electrical
connection (see the figure below).
F0328.ai
(4) Rotate the transmitter or the terminal box in the
desired direction. The direction can be changed to -90
degrees, +90 degrees or +180 degrees.
NOTE
The terminal box can be rotated -90 degrees to +180
degrees at every 90 degrees from the arrow mark
indicating the flow direction. Do not exceed these
angle.
(5) Tighten diagonally the four hexagonal bolts with the
hexagonal wrench, and then be sure to check that the
terminal box was firmly fixed to the sensor.
F0329.ai
AXW Remote Sensor
±0°
+90°
+180°
-90°