![YOKOGAWA ZR22S User Manual Download Page 188](http://html1.mh-extra.com/html/yokogawa/zr22s/zr22s_user-manual_3505092188.webp)
<12. Troubleshooting>
12-12
12.3.2 Measured Value Lower Than True Value
<Causes and Countermeasures>
(1) The sample gas pressure becomes lower.
Where an increment of the measured value due to pressure change cannot be neglected,
take measures referring to Subsection 12.3.1 (1).
(2) Moisture content in a reference gas changes (decreases) greatly.
If air at the detector installation site is used for the reference gas, large change of moisture
content in the air may cause an error in measured oxygen concentration value (vol%O
2
).
When this error is not ignored, use a gas in which moisture content is constant such as
instrument air in almost dry condition as a reference gas.
In addition, change of moisture content in exhaust gas after combustion is also considered
as a cause of error. However, normally this error is negligible.
(3) Calibration gas (zero gas) is mixed into the detector due to leakage.
If the zero gas is mixed into the detector due to leakage as a result of failure of the valve
provided in the calibration gas tubing system, the measured value shows a value a little
lower than normal.
Check valves (needle valves, check valves, solenoid valves for automatic calibration,
etc.) in the calibration gas tubing system for leakage. For manual valves, check them after
confirming that they are in a fully closed situation.
(4) Combustible components exist in the sample gas. If combustible components exist in the
sample gas, they burn in the sensor and thus oxygen concentration decreases.
Check that there are no combustible components.
(5) Temperature of the detector cell reaches 750°C or more.
12.3.3 Measurements Sometimes Show Abnormal Values
<Cause and Countermeasures>
(1) Noise may be mixing in with the converter from the detector output wiring.
Check whether the converter and detector are securely grounded.
Check whether or not the signal wiring is laid along other power cords.
(2) The converter may be affected by noise from the power supply.
Check whether or not the converter power is supplied from the same outlet, switch, or
breaker as other power machines and equipment.
(3) Poor wiring contact.
If there is poor contact in the wiring, the sensor voltage or thermocouple e.m.f. (voltage) may
vary due to vibration or other factors.
Check whether or not there are loose points in the wiring connections or loose crimping
(caulking) at the crimp-on terminal lugs.
(4) There may be a crack in the sensor or leakage at the sensor-mounting portion.
If the indication of concentration varies in synchronization with the pressure change in the
furnace, check whether or not there is a crack in the sensor or whether the sensor flange is
sticking tightly to the probe-attaching face with the metal O-ring squeezed.
(5) There may be leakage in the calibration gas tubing.
In the case of a negative furnace inner pressure, if the indication of concentration varies with
the pressure change in the furnace, check whether or not there is leakage in the calibration
gas tubing.
IM 11M13A01-02E
Summary of Contents for ZR22S
Page 11: ...Blank Page...
Page 19: ...Blank Page...
Page 23: ...Blank Page...
Page 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Page 70: ...Blank Page...
Page 134: ...Blank Page...
Page 164: ...Blank Page...
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Blank Page...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......