<8. Detailed Data Setting>
8-2
IM 11M13A01-04E
8.2.1 Setting Minimum Oxygen Concentration ( at 4 mA) and Maximum Oxygen
Concentration ( at 20 mA)
Set the oxygen concentration values at 4 mA and 20 mA.
The minimum concentration of oxygen for the minimum current (4 mA) is 0%O
2
or 6% to 76%O
2
.
The maximum concentration of oxygen for the maximum current (20 mA) ranges from 5% to 100%
O
2
, and must be greater than 1.3 times the concentration of oxygen set for the minimum. If it does not
fall within this input range setting, the setting will be invalid, and the previous set values will remain.
Setting example 1
If the setting (for a 4 mA current) is 10%O
2
, you must set the oxygen concentration for the maximum
(20 mA) point at 13%O
2
.
Setting example 2
If the setting (for a 4 mA current) is 75%O
2
, you must set the oxygen concentration for the maximum
(20 mA) point at 98%O
2
(75 × 1.3).
(Numbers after the decimal point are rounded up.)
CAUTION
• When you select logarithmic mode, the minimum output remains constant at 0.1%O
2
, and the
parameter [C11] display remains unchanged.
8.2.2 Entering Output Damping Constants
If a measured value adversely affected by a rapid change in the sample gas oxygen concentration is
used for the control means, frequent on-off actions of the output will result. To avoid this, the analyzer
allows the setting of output damping constants ranging from 0 to 255 seconds.
8.2.3 Selection of Output Mode
You can select a linear or logarithmic output mode. The former provides linear characteristics between
the analog output signal and oxygen concentration.
8.2.4 Default Values
When the analyzer is delivered or data are initialized, the current output settings are by default as
shown in Table 8.3.
Table 8.3
Current Output Default Values
Item
Default setting
Min. oxygen concentration
0%
O
2
Max. oxygen concentration
25%
O
2
Output damping constant
0 (seconds)
Output mode
Linear
CAUTION
• When you select logarithmic mode, the minimum output remains constant at 0.1%O
2
, and the
parameter [C11] display remains unchanged.
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......