
<9. Calibration>
9-2
IIM 11M13A01-04E
0.1
0.5
1
5
10
21.0
50
100
120
100
80
60
40
20
0
-20
-40
0.51 vol%O
2
,81.92mV(Zero origin of calibration)
21.0 vol%O
2
, 0mV
(Span origin of calibration)
Oxygen concentration (vol % O
2
)
Cell
voltage
(mV)
F9-1E.ai
Figure 9.1
Oxygen Concentration in a Measurement Gas vs. Cell Voltage
(21 vol%O
2
Equivalent)
The measurement principles of a zirconia oxygen analyzer have been described above. However, the
relationship between oxygen concentration and the electromotive force of a cell is only theoretical.
Usually, in practice, a sensor shows a slight deviation from the theoretical value. This is the reason
why calibration is necessary. To meet this requirement, an analyzer calibration is conducted so that a
calibration curve is obtained, which corrects the deviation from the theoretical cell electromotive force.
9.1.2 Calibration Gas
A gas with a known oxygen concentration is used for calibration. Normal calibration is performed
using two different gases: a zero gas of low oxygen concentration and a span gas of high oxygen
concentration. In some cases, only one of the gases needs to be used for calibration. However, even if
only one of the gases is normally used, calibration using both gases should be done at least once.
The zero gas normally used has an oxygen concentration of 0.95 to 1.0 vol%O
2
with a balance of
nitrogen gas (N
2
). The span gas widely used is clean air (at a dew-point temperature below -20°C and
free of oily mist or dust, as in instrument air).
Summary of Contents for ZR202S
Page 12: ...xi IM 11M13A01 04E Introduction ...
Page 18: ...Blank Page ...
Page 61: ...Blank Page ...
Page 129: ...Blank Page ...
Page 141: ...Blank Page ...
Page 153: ......
Page 155: ......
Page 157: ......
Page 159: ...Blank Page ...
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......