2-5
IM WT5000-02EN
2.2 Setting the Voltage Range and Current Range
Setting the Current Range (Current Range) When the Input
Element Is 760901 or 760902
3.
Tap
Current Range
of the input element number you want to configure. A current range setup
menu appears.
4.
Follow the instructions below to set the current range.
• Sets auto range mode
Tap
Auto
. The AUTO key illuminates.
• Setting the Fixed Range
Tap
Current Range
. A current range menu appears.
Tap a current range on the menu to set the current range.
Set the current range to auto.
Select the current range.
Current range (example of element 1)
Turns the external current sensor on or off (see section 2.3).
Configure the current sensor (see section 2.3).
Operations in the Input Information Area
See page 2-8.
Available current range options
• 30 A Input Element 760901
When the Crest Factor Is Set to CF3
When the Crest Factor Is Set to CF6 or CF6A
500 mA, 1 A, 2 A, 5 A, 10 A, 20 A, 30 A
250 mA, 500 mA, 1 A, 2.5 A, 5 A, 10 A, 15 A
• 5 A Input Element 760902
When the Crest Factor Is Set to CF3
When the Crest Factor Is Set to CF6 or CF6A
5 mA, 10 mA, 20 mA, 50 mA, 100 mA,
200 mA, 500 mA, 1 A, 2 A, 5 A
2.5 mA, 5 mA, 10 mA, 25 mA, 50 mA, 100 mA,
250 mA, 500 mA, 1 A, 2.5 A
Note
When range Σ link (see section 2.1) is set to on, the current ranges of the input elements assigned to the
same wiring unit are set to the same range. When range Σ link is set to off, the current ranges of the input
elements can be set independently even when they are assigned to the same wiring unit.