![YOKOGAWA WT5000 User Manual Download Page 142](http://html.mh-extra.com/html/yokogawa/wt5000/wt5000_user-manual_911423142.webp)
7-4
IM WT5000-02EN
• When the Storage Mode Is Integ Sync (integration synchronization)
Computation/Output tab
Configure the storage.
Storage control setup screen (when the storage mode is Integ Sync)
Set the storage interval
(see page 7-2).
Set whether to save (store) numeric data
at the start of storage.
OFF: Not be saved
ON: Saved
You cannot set this when the storage interval is set to 00:00:00.
• When the Storage Mode Is Event
Computation/Output tab
Configure the storage.
Storage control setup screen (when the storage mode is Event)
Select the trigger event (Ev1 to Ev8).
When measured data is updated, storage is started
if the conditions of the specified user-defined event
are met.
Note
When the storage mode is set to Single Shot, there are no storage control settings.
7.1 Setting the Storage Operation