7-1
IM WT5000-02EN
Chapter 7 Storing Numeric Data
7.1 Setting the Storage Operation
► “Data Storage (Store, MENU (STORE))” in the features guide
► “Storage Control” in the features guide
This section explains operating procedures using the following setup methods.
• Procedure Using the Setup Menu (see chapter 1)
• Procedure Using the Menu Icons (see page iii)
• Procedure Using the Keys (other than SETUP) (see section 1.2 in IM WT5000-03EN)
Procedure Using the Setup Menu
1.
Tap the
Setup
icon
, or press
MENU
under SETUP.
2.
Tap
Computation/Output
tab. A computation and output settings overview screen appears.
Pressing
ESC
closes the overview screen.
Configuring the Storage (Store)
3.
Tap
Store
.
A Store screen appears.
Computation/Output tab
Configure the
storage.
Set the storage mode (Manual, Real Time, Integ Sync, Event, Single Shot).
Set the storage
count (Infinite, 1 to
9999999)
Sets the storage count to its
maximum
You can set the storage count to its
maximum, which is calculated
automatically from the save
destination space and the amount
of numeric data to be saved.
See section 7.2.
See section 7.3.
See “Configuring the Storage
Control (Control Settings)” on
the next page.
Note
You can also display the computation/output settings overview screen by moving the cursor on the
Computation/Output tab using the arrow keys and then pressing SET.