![YOKOGAWA WT500 User Manual Download Page 292](http://html.mh-extra.com/html/yokogawa/wt500/wt500_user-manual_911418292.webp)
13-1
IM 760201-01E
Troubleshooting, Maintenance, and Inspection
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
13.1 Troubleshooting
Dealing with Unusual Circumstances
• If a message appears on the screen, see the following pages for reference.
• If servicing is necessary, or if the instrument does not operate properly even after you
have attempted to deal with the problem according to the instructions in this section,
contact your nearest YOKOGAWA dealer.
Problems and Solutions
Refer to
section
Nothing appears on the screen when you turn on the power switch.
Securely connect the power cord to the instrument and to the power outlet.
3.3
Set the supply voltage to within the permitted range.
3.3
The displayed data is not correct.
Check that the ambient temperature and humidity are within their specified
ranges.
3.2
Check to make sure that noise is not affecting the measurement.
3.1, 3.5
Check the measurement cable wiring.
3.8 to 3.11
Check the wiring system.
3.8 to 3.11,
4.2
Check that the line filter is off.
4.8
Check the measurement period settings.
4.7
Check the FAQ at the following URL.
http://www.yokogawa.com/tm/
—
Turn the power off and then on again.
3.4
Keys are not functioning.
Check the REMOTE indicator. If the REMOTE indicator is lighted, press
LOCAL to turn it off.
—
Check that the key lock is off.
12.5
Perform a key test. If the test fails, servicing is necessary.
13.3
Triggering does not work.
Check the trigger conditions.
7.5
Check that the trigger source is being applied.
7.5
Unable to make harmonic
measurements.
Check the PLL source settings.
6.3
Check that the PLL source input signal is within the specifications.
6.3
Unable to recognize storage medium. The storage medium may be damaged.
—
Unable to save data to the selected medium.
Check the free space on the medium. Remove files or use a different medium
as necessary.
10.2
Unable to configure or control the instrument through the communication interface.
Check that the communication interface parameters, such as the serial
number (for USB), GP-IB address, and IP address (for Ethernet), meet the
specifications.
—
*
Check that the interface meets the electrical and mechanical specifications.
—
*
* See the
Communication Interface User’s Manual (IM760201-17E)
.
Chapter 13 Troubleshooting, Maintenance, and Inspection