5-21
IM 760201-01E
Power Measurement
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Setting the Integration Timer
8.
Use the
cursor keys
to select one of the Integ Timer boxes (hour, minute, or
second).
9.
Press
SET
to display an entry box.
10.
Use the
cursor keys
to set the hour, minute, or second that you selected in step
8 to zero.
11.
Press
SET
or
ESC
to close the entry box.
12.
Repeat steps 8 to 11 to set the hour, minute, and second to 00000:00:00.
13.
Press
ESC
twice to close the entry box and the setup menu.
Integrating in Manual Integration Mode
• Starting Integration
14.
Press
START/STOP
. The integration indicator changes to Start (displayed in
green), and integration starts.
• Holding Integration
15.
Press
HOLD
. The HOLD key lights, and the numeric data display is held. The
WT500 continues the integration operation internally.
• Un-Holding
16.
Press
HOLD
when the display is held. The HOLD key is no longer lighted, and
the numeric data display is updated. While the display is held, you can execute a
single measurement by pressing SHIFT+HOLD (SINGLE) to update the display.
• Stopping Integration
17.
Press
START/STOP
. The integration indicator changes to Stop (displayed in red).
The integration time and integrated value are held. You can press
START/STOP
again to resume integration.
• Resetting Integration
18.
Press
SHIFT+START/STOP (RESET)
. The integration indicator changes to
Reset, the integration time and integrated value are reset, and the displays for
integration-related functions such as WP change to “-----------” (no data).
5.7 Manual Integration