![YOKOGAWA WT500 User Manual Download Page 141](http://html.mh-extra.com/html/yokogawa/wt500/wt500_user-manual_911418141.webp)
5-15
IM 760201-01E
Power Measurement
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
5.6 Integration
The integration feature has the following five modes.
Integration Mode
Manual integration mode
A WT500 key is used to start/stop
integration.
Normal integration mode
A WT500 key is used to start
integration. Integration ends when a
specified amount of time elapses.
Continuous integration mode
A WT500 key is used to start
integration. Integration resets when
a specified amount of time elapses.
Integration repeats until you press
STOP.
Real-time normal integration mode
The integration start and end times
are specified.
Real-time continuous integration mode
The integration start and end times
are specified. Integration resets
when a specified amount of time
elapses.
Start
Key
operation
Key
operation
Key
operation
Date and
time
Date and
time
Stop
Key
operation
Stopped
by the timer
Key
operation
Date and
time
Date and
time
---
---
Repetition
Repeats
when the timer
expires
---
Repeats
when the timer
expires
Reference
Section
5.7
Section
5.8
Section
5.8
Section
5.9
Section
5.9
For details on each integration mode, see section 2.6.
Starting, Stopping, and Resetting Integration
• You can start, stop, or reset integration by using the front panel keys or communication
commands.
• The following figure shows the relationship between the integration operation and
the actions of starting, stopping, and resetting. Pressing RESET after integration has
stopped resets the integrated value and the integration time.
Integrated value
Integration time
Abort
Start
Restart
Abort
Reset
Restart
Auto stop
Reset
Timer value
START STOP
START
STOP
RESET
START