1-15
IM WT310E-01EN
Features
1
2
3
4
5
6
7
8
App
Index
1.5
Integrated Power (Watt hour)
You can integrate the active power (watt hour) and current (ampere hour).
For information about the measurement function symbols for integrated power (watt hour) and their
meanings, see “Items That This Instrument Can Measure.”
Integration Modes (For procedures, see section 5.2)
The integration feature has the following three modes.
Integration Mode
Start
Stop
Repetition
Manual integration mode
Key operation
Key operation
---
Standard integration mode
Key operation
Stopped by the timer
---
Continuous (repetitive) integration
mode
Key operation
Key operation
Repeats when the timer
expires
Manual Integration Mode
When you set the integration mode to standard integration mode (nor) and set the integration timer
to 00000:00:00, this instrument performs integration in manual integration mode. When you press
START, integration starts, and it continues until you press STOP. However, if either of the conditions
below is met, integration is stopped, and the integration time and integrated value are held.
• The integration time reaches the maximum integration time (10000 hours).
• The integrated value reaches its maximum or minimum displayable value.
Reset
Start
Stop
Start
Reset
Hold
Hold
Hold
Hold
When the maximum
integrated value is reached
Integrated
value
Integration
time