7.13 General Specifications
Item
Specifications
Warm-up time
Approx. 30 minutes
Operating environment
Temperature: 5°C to 40°C
Humidity: 20%RH to 80%RH (no condensation)
Elevation: 2000 m or less
Installation location
Indoors
Storage environment
Temperature: -25°C to 60°C
Humidity: 20%RH to 80%RH (no condensation)
Rated supply voltage
100 VAC to 240 VAC
Permitted supply voltage
range
90 VAC to 264 VAC
Rated supply frequency
50/60 Hz
Permitted supply voltage
frequency range
48 Hz to 63 Hz
Maximum power
consumption
WT310, WT310HC: 50 VA, WT332/WT333: 70 VA
External dimensions
(excluding protrusions)
WT310, WT310HC: Approx. 213 (W) × 88 (H) × 379 (D) mm
WT332/WT333: Approx. 213 (W) × 132 (H) × 379 (D) mm
Weight
WT310, WT310HC: Approx. 3 kg
WT332/WT333: Approx. 5 kg
Battery backup
Setup parameters are backed up with a lithium battery.
Standard accessories
• Power cord: 1 piece
• D/A cable: 1 piece (included in products with the /DA4, /DA12 options)
• Current input protection cover: 1 piece that matches the main unit
• Hind feet rubber: 1 set
• User’s manual: 1 piece (this manual)
• Safety terminal adapter:
WT310, WT310HC: Red/Black 1 set
WT332: Red/Black 2 sets
WT333: Red/Black 3 sets
Safety standard
1
Compliant standard: EN61010-1
Installation category (overvoltage category) CAT II
2
Measurement Category CAT II
3
Pollution degree 2
4
Emission
1
Compliant standards
EN61326-1 Class A
EN55011 ClassA, Group1
EN61000-3-2
EN61000-3-3
C-tick EN55011 Class A, Group1
Korea Electromagnetic Conformity Standard (
한국 전자파적합성기준
)
This product is a Class A (for industrial environment) product. Operation of this product in a
residential area may cause radio interference in which case the user will be required to correct
the interference.
Cable conditions
• Serial (RS-232) interface connector
Use shielded serial cables.
5
• GP-IB interface connector
Use shielded GP-IB cables.
5
• USB port (PC)
Use shielded USB cables.
5
• Ethernet port
Use category 5 or better Ethernet cables (STP).
6
• D/A output terminal
Use shielded cables.
5
• Connecting cables may cause radio interference in which case the user is required to
correct the interference.
Immunity
1
Compliant standard
EN61326-1 Table 2 (for industrial locations)
However, the measured values may receive noise effects.
Cable conditions
Same as the cable conditions for emission above.