9.3 Selecting the IEC Harmonic Measurement Mode
Procedure
START STOP
REMOTE
CAL
SENSOR RATIO
MEASURING
RMS
MEAN
DC
RMEAN
CURRENT RANGE
VOLTAGE RANGE
MEASURING
RMS
MEAN
DC
RMEAN
ALL
INPUT INFO.
LOWER ITEM
LOWER FORM
ALL
USER SET
DISPLAY
ITEM & ELEMENT
2
3
4
COMPEN
ELEMENT
ELEMENT
1
WIRING
MODE
AUTO
MODE
AUTO
EXT SENSOR
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
RESET
SET
PAGE
PAGE
Cursor keys
To exit the menu during
operation, press
ESC
located
above the soft keys.
In the procedural explanation
below, the phrase “press the
cursor keys
” may be used.
This phrase refers to the
procedures for selecting
items and entering values
and character strings. For
details on the procedures,
see section 3.14 in the
User’s
Manual
, IM WT3001E-01EN
.
• Enabling IEC Harmonic Measurement Mode
1.
Press
OTHERS
to display the Others menu.
2.
Press the
IEC Harmonics
soft key.
Pres the Next1/3
soft key
Explanation
If you use the Harmonic/Flicker Measurement Software, this instrument automatically
switches to IEC harmonic measurement mode.
This instrument cannot switch to IEC harmonic measurement mode in the following
conditions.
• When the master/slave synchronization measurement is set to slave.
• When the integration is started or stopped.
• When auto print is ON.
• While storage or recall operation is in progress.
• While the storage medium is being accessed.
• While saving waveform sampling data to the storage medium.