6-88
IM WT3001E-17EN
:NUMeric:FLICker:INFOrmation:CLEar
Function
Clears (sets to NONE) the output items of flicker
judgment results (variable format).
Syntax
:NUMeric:FLICker:INFOrmation:
CLEar {ALL|<NRf>[,<NRf>]}
ALL = Clears all items
1st <NRf> = 1 to 32 (item number to start
clearing)
2nd <NRf> = 1 to 32 (item number to stop
clearing)
Example
:NUMERIC:FLICKER:INFORMATION:
CLEAR ALL
Description • Only available with the flicker measurement
function (/FL option).
• If the 2nd <NRf> is omitted, the output items
from the start clear number to the last item (32)
are cleared.
:NUMeric:FLICker:INFOrmation:DELete
Function
Deletes the output items of flicker judgment
results (variable format).
Syntax
:NUMeric:FLICker:INFOrmation:
DELete {<NRf>[,<NRf>]}
1st <NRf> = 1 to 32 (item number to start
deleting)
2nd <NRf> = 1 to 32 (item number to stop
deleting)
Example
:NUMERIC:FLICKER:INFORMATION:
DELETE 1
(Deletes ITEM1 and shifts ITEM2 and subsequent
items forward)
:NUMERIC:FLICKER:INFORMATION:
DELETE 1,3
(Deletes ITEM1Å|3 and shifts ITEM4 and
subsequent items forward)
Description • Only available with the flicker measurement
function (/FL option).
• Output items subsequent to the deleted output
items are shifted in order into the deleted items’
position, and NONE is set in the open positions
on the end.
• If the 2nd <NRf> is omitted, only the output
item of the start delete number is deleted.
:NUMeric:FLICker:INFOrmation:ITEM<x>
Function
Sets the output items (function, element, and
observation period) of flicker judgment results
(variable format) or queries the current setting.
Syntax
:NUMeric:FLICker:INFOrmation:
ITEM<x> {NONE|<Function>,
<Element>[,<Period>]}
:NUMeric:FLICker:INFOrmation:
ITEM<x>?
<x> = 1 to 32 (item number)
NONE = No output items
<Function> =
{JTOTal|JDC|JDMAX|JDT|
JTMAX|JPST|JPLT}
<Element> =
{<NRf>|TOTal}
(<NRf> = 1 to 4)
<Period> =
{<NRf>|ALL}
(<NRf> = 1 to 99)
Example
:NUMERIC:FLICKER:INFORMATION:
ITEM1 JDC,1,1
:NUMERIC:FLICKER:INFORMATION:
ITEM1? -> :NUMERIC:FLICKER:
INFORMATION:ITEM1 JDC,1,1
Description • Only available with the flicker measurement
function (/FL option).
• The contents that are output for each of the
selections for <Function> are as follows:
<Function>
Output Contents
<Element>
Designation
<Period>
Designation
JTOTal
Overall judgment results for dc,
dmax, d(t)*, pst, and plt
Required
Not required
(the judgment results displayed under [Element# Judgment:]
in the upper right part of the screen)
JDC
Judgment results for relative
steady-state voltage change dc
Required
Required
JDMAX
Judgment results for maximum
relative voltage change dmax
Required
Required
JDT
Judgment results for relative
voltage change time d(t)
Required
Required
JTMAX
Judgment results for Tmax
Required
Required
JPST
Judgment results for short-term
flicker value Pst
Required
Required
JPLT
Judgment results for long-term
flicker value Plt
Required
Not required
* Tmax for IEC 61000-3-3 Edition 3.0, d(t) for
IEC 61000-3-3 Edition 2.0.
6.18 NUMeric Group