Faunctional Description
2.8
Trend, Bar Graph, and Vector Displays
Trends of each measurement function, bar graphs of harmonics of each order, and
vectors of the fundamental signal of each element (when using the harmonic
measurement option) can be displayed.
Trend display
The trends of all measurement functions that are measured during normal measurement
and harmonic measurement are displayed.
Trend of voltage U1
Trend of the
frequency of
voltage U1
• Trend Display Data
When the retrieval of waveform display data is OFF during normal measurement, the
numeric data of measurement functions that are determined for each data update rate
is P-P compressed
*
for each display segment (raster) and made into trend display
data.
When the retrieval of waveform display data is ON during normal measurement, the
numeric data of measurement functions that are determined each time a trigger
occurs is P-P compressed
*
for each display segment (raster) and made into trend
display data.
*
In some cases, P-P compression is not performed.
• Horizontal (Time) Axis «For procedures, see section 7.4.»
The time per division can be set in the range of 3 s to 1 day.
• Setting the Scale «For procedures, see section 7.5.»
Auto scaling is provided in which the upper and lower limits of the screen are
determined automatically from the maximum and minimum values of the trend display
data. Manual scaling is also available in which the upper and lower limits can be set
arbitrarily as necessary.
• Split Display Display and Assignments «For procedures, see section 7.6.»
Up to 16 lines (T1 through T16) of trends can be displayed. You can select the
measurement function of any element to be assigned to T1 through T16. During
harmonic measurement, you can also specify the harmonic.
The screen is divided up to 4 windows and the trends that are turned ON are
assigned to the divided windows in order from T1 through T16.
• Display Interpolation, Graticule, and Label Display «For procedures, see
sections 6.6 and 6.7.»
The settings specified for waveform display are used.