2-13
IM 760301-01E
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Faunctional Description
Averaging «For procedures, see section 4.9.»
The averaging function is effective when reading of the numeric display is difficult due to
fluctuations. This occurs when the fluctuation of the power supply or the load is large or
when the input signal frequency is low.
• Measurement Functions of Normal Measurement
Two types, exponential average and moving average, are available.
• Exponential Average
The numeric data can be exponentially averaged using a specified attenuation
constant. Averaging is performed according to the following equation.
D
n
=
( M
n
– D
n – 1
)
K
D
n – 1 +
D
n
:
Displayed value that has been exponentially averaged n times. (The
displayed value D
1
at the first time is M
1
.)
D
n–1
: Displayed value that has been exponentially averaged n–1 times.
M
n
:
Measured data at the n
th
time.
K:
Attenuation constant (select from 2, 4, 8, 16, 32, and 64)
• Moving Average
The numeric data can be linearly averaged using a specified average count.
Averaging is performed according to the following equation.
D
n
=
M
n – ( m – 1 )
+ · · · M
n – 2
+ M
n – 1
+ M
n
m
D
n
: Displayed value obtained by linearly averaging m points of numeric data from
the n–(m–1)
th
to n
th
time
M
n–(m–1)
: Measured data at the n–(m–1)
th
time.
••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••
M
n–2
: Measured data at the (n–2)
th
time.
M
n–1
: Measured data at the (n–1)
th
time.
M
n
:
Measured data at the n
th
time.
m:
Average count (select from 8, 16, 32, 64, 128, and 256)
• Measurement Functions of Harmonic Measurement
• Harmonic Measurement in Normal Measurement Mode
When averaging is turned ON, averaging is performed on the measurement
functions of harmonic measurement when exponential averaging (above) is
specified.
If averaging is set to moving average, averaging is performed only on the
measurement functions of normal measurement and not the measurement
functions of harmonic measurement.
• Measurement Functions of Wide Bandwidth Harmonic Measurement Mode
Performs exponential averaging. The attenuation constant can be selected from
above.
• Measurement Functions of IEC Harmonic Measurement Mode
Performs exponential averaging. The attenuation constant cannot be selected.
• Measurement Functions of Voltage Fluctuation and Flicker Measurement Mode
and Cycle-by-Cycle Measurement Mode
Averaging is not performed. The average setting is invalid.
2.3 Measurement Conditions