10.2 NULL Function
Procedure
START
STOP
REMOTE
CAL
MOTOR
SET
CURSOR
NULL
MENU
STORE SET
MENU
LOWER ITEM
LOWER FORM
ALL
USER SET
DISPLAY
ITEM & ELEMENT
RESET
SET
PAGE
PAGE
SCALING
HRM SET
MEASURE
SYNC SOURCE
AVG
LINE FILTER
FREQ
FILTER
FILE
MISC
STORE
IMAGE SAVE
HELP
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
To exit the menu during operation, press
ESC
located above
the soft keys.
In the procedural explanation below, the phrase “press the
cursor keys
” may be used. This phrase refers to the
procedures for selecting items and entering values and
character strings. For details on the procedures, see section
3.14.
Cursor keys
• Activating the NULL Function
1.
Press
SHIFT
+
SYNC SOURCE(NULL)
. The NULL indicator to the lower left of the key
illuminates when the NULL function operates and the measured values are updated.
• Clearing the NULL Function
2.
Press
SHIFT
+
SYNC SOURCE (NULL)
when the NULL function is activated.
The NULL indicator turns OFF and the NULL function is released.
Explanation
The NULL function can be used to subtract the DC offset while the external sensor or
measurement cable is connected. Execute the NULL function at no input condition
(opened current input terminal and shorted voltage input terminal).
• NULL Value
The NULL values are set to the Udc and Idc values (numeric data of the simple
average of voltage/current) as well as speed and torque (on models with the motor
evaluation function when the input signal from the sensor is analog) that were
measured the previous time when the voltage/current mode is set to DC and the
NULL function is turned ON. If the previously measured Udc and Idc as well as Speed
and Torque values do not exist, such as when the NULL function is turned ON without
making a measurement after turning ON the power, Udc and Idc are set to zero. To
subtract the DC offset using the NULL function when making measurements with the
voltage/current mode set to RMS, MEAN, or RMEAN, activate the NULL function with
the voltage/current mode set to DC and then change the voltage/current mode to
RMS, MEAN, or RMEAN.
Note
•
The NULL function is automatically released when the range or filter setting is changed or
integration is started.
•
If you perform the procedure of releasing the NULL function while in the hold condition, the
NULL function is actually released and the NULL indicator turns OFF when the hold is
released or when a single measurement is made.
•
If the crest factor is set to 3, a NULL value of approximately 50% is set for a signal that
exceeds 50% of the range. If the crest factor is set to 6, a NULL value of approximately 100%
is set for a signal that exceeds 100% of the range.