ii
IM 760401-01E
Functional Comparison of the WT210/WT230 and
WT200/WT110E/WT130
This section summarizes the functional differences between the WT210/WT230 and the
conventional models WT200/WT110E/WT130. For more details on the functions and
performance of the WT210/WT230, see the following sections.
Item
WT210/WT230
WT200/WT110E/WT130
(Conventional Models)
Voltage input terminal
Plug-in terminal
Binding post
(safety terminal structure)
External sensor input terminal
BNC connector
Plug-in terminal (safety terminal
structure)
Basic voltage and
0.1% of reading
0.15% of reading
current accuracy
+0.1% of range
+0.1% of range
Basic power accuracy
0.1% of reading
0.2% of reading
+0.1% of range
+0.1% of range
Frequency range
DC, 0.5 Hz to 100 kHz
DC, 10 Hz to 20 kHz
Effective input range
1 to 130% of the rated range
10 to 130% of the rated range
Measurement synchronization Select from “voltage,” “current,” WT200: Select voltage or current
source
and “entire period of the display WT110E/WT130: No
update rate.”
Line filter
Yes (cutoff frequency 500 Hz)
None
Frequency filter
Yes (cutoff frequency 500 Hz)
Yes (cutoff frequency 300 Hz)
MAX hold
Yes
WT200: Yes
WT110E/WT130: No
Peak value display
Yes
WT200: Yes, WT110E: Option
WT130: Yes
Average active power during
Yes
WT200: Yes
integration
WT110E/WT130: No
Display update rate
Select 0.1, 0.25, 0.5, 1, 2,
Fixed to 0.25 s
or 5 s
Number of displayed digits
Select 4 or 5 digits
WT200: Select 4 or 5 digits
WT110E/WT130: Fixed to 4
digits
Integration timer time
10000 hours maximum
WT200:
Resolution 1 s
10000 hours maximum
Resolution 1 s
WT110E/WT130:
999 hours 59 minutes maximum
Resolution 1 min
Display update during
Select 0.25, 0.5, 1, 2, or 5 s
Approx. 3 s
harmonic measurement
Remote control I/O signal when
EXT HOLD, EXT TRIG,
EXT HOLD, EXT TRIG
equipped with the comparator
EXT START, EXT STOP,
function (/CMP option)
EXT RESET, INTEG BUSY
Communication Commands
All communication commands for conventional models can be
used except the commands for the 2533E.
Communication data format
ASCII and binary
ASCII
Addressable mode B of GP-IB No
Yes
communications
Baud rate of serial (RS-232-C) 1200 to 9600 bps
75 to 9600 bps
communications
Zero-level compensation
Yes
WT200: Yes, WT110E/WT130: No
Key lock
Yes
No
Power fuse
Yes, part number A1347EF
WT200: No, WT110E: No
WT130: Yes, part number
A1346EF
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...