Specifications
16
16.3 Functions
Voltage, Current, and Active Power Measurements
Item
Specifications
Measurement method
Digital sampling method, summation averaging method
Crest factor
3 or 6
Wiring system
WT210 (single-phase model, 760401)
Single-phase, two-wire (1P2W)
WT230 (three-phase, three-wire model, 760502)
Select from single-phase, two-wire (1P2W); single-phase, three-wire (1P3W); and three-phase,
three-wire (3P3W)
WT230 (three-phase, four-wire model, 760503)
Select from single-phase, two-wire (1P2W); single-phase, three-wire (1P3W); three-phase,
three-wire (3P3W); three-phase, four-wire (3P4W); and three-voltage, three-current (3V3A).
Range switching
Select fixed range or auto range.
Auto range
Step up
When the measured value exceeds 130% of the rated range or the peak value exceeds
approximately 300% (Approx. 600% if the crest factor is set to 6) of the rated range. For the
WT230, when the condition is met on any of the input elements.
Step down
Range down: When the measured value becomes less than 30% of the rated range and the
peak value is less than 300% (Less than or equal to 600% if the crest factor is set to 6) of the
subordinate range. For the WT230, when the condition is met on all input elements.
Measurement mode
Select from RMS (the true RMS value of voltage and current), VOLTAGE MEAN (the rectified mean
switching
value calibrated to the RMS value of the voltage and the true RMS value of the current), DC (simple
average of voltage and current).
Measurement
Select voltage, current, or the entire period of the display update rate for the signal used to achieve
synchronization source
synchronization during measurement.
Line filter
Select OFF or ON (cutoff frequency of 500 Hz).
Peak measurement
Measures the peak value of voltage or current from the instantaneous voltage or instantaneous
current that is sampled.
Zero-level compensation
Creates a zero input condition inside the WT210/WT230 and sets the level at that point as the zero
level.
Frequency Measurement
Item
Specifications
Measured item
WT210 (single-phase model, 760401)
Select voltage (V) or current (A).
WT230 (three-phase, three-wire model, 760502)
Select voltage of input element 1 (V1), voltage of input element 3 (V3), current of input element 1
(A1), or current of input element 3 (A3).
WT230 (three-phase, four-wire model, 760503)
Select voltage of input element 1 (V1), voltage of input element 2 (V2), voltage of input element 3
(V3), current of input element 1 (A1), current of input element 2 (A2), or current of input element
3 (A3).
Method
Reciprocal method
Frequency measuring range
Varies depending on the display update period (see description given later) as follows.
Display Update Rate
Measurement Range
0.1 s
25 Hz to 100 kHz
0.25 s
10 Hz to 100 kHz
0.5 s
5 Hz to 100 kHz
1 s
2.5 Hz to 100 kHz
2 s
1.5 Hz to 50 kHz
5 s
0.5 Hz to 20 kHz
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...