13-14
IM 760401-01E
• You cannot change the setting while
integration is in progress. Execution error 13
will occur.
• While recalling or storing is in progress,
execution error 19 will occur.
SL
Recalls setup parameters from a
selected file.
Syntax
SL m <terminator>
“m” indicates file no., and must be
set within the following range.
1
≤
m
≤
4
Description • Parameter error 12 will occur if “m” is set to
an illegal value.
• It is not possible to recall communications-
related information (communication mode,
address etc.) using this command.
• While recalling or storing is in progress,
execution error 19 will occur.
SN/SN?
Sets the measurement synchronization
source/inquires about the current
setting.
Syntax
SN m <terminator>
“m” indicates the type of
measurement synchronization source.
m= 0: V (voltage)
1: A (current)
2: OFF
Query
SN? <terminator>
Example
SN0
Description • Parameter error 12 will occur if “m” is set to
an illegal value.
• You cannot change the setting while
integration is in progress. Execution error 13
will occur.
• While recalling or storing is in progress,
execution error 19 will occur.
SO/SO?
Sets the store function ON/OFF or
inquires about the current setting.
Syntax
SO m <terminator>
“m” indicates whether storage is ON
or OFF.
m= 0: OFF
1: ON
Query
SO? <terminator>
Example
SO1
Description • Parameter error 12 will occur if “m” is set to
an illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
SR/SR?
Sets the storage interval/inquires about
the current setting.
Syntax
SR m1,m2,m3 <terminator>
“m1” indicates the hours
0
≤
m1
≤
99
“m2” indicates the minutes
0
≤
m2
≤
59
13.1 Commands
“m3” indicates the seconds
0
≤
m3
≤
59
Query
SR? <terminator>
Example
SR0,0,0
Description • Parameter error 12 will occur if an illegal
value is set.
• If the store interval is set to 0 h 0 min 0 s, the
store interval is set to the same interval as the
display update rate.
• While recalling or storing is in progress,
execution error 19 will occur.
SS
Stores setup parameters into a selected
file.
Syntax
SS m <terminator>
“m” indicates file no., and must be
set within the following range.
1
≤
m
≤
4
Description • Parameter error 12 will occur if “m” is set to
an illegal value.
• The following setup parameters can be
stored:
All setup parameters which can be output by
the OS command
Information related to communications (GP-
IB, serial, etc.)
TM/TM?
Sets integration preset time/inquires
about the current setting.
Syntax
TM m1,m2, m3 <terminator>
“m1” indicates the hours
0
≤
m1
≤
1000
“m2” indicates the minutes
0
≤
m2
≤
59
“m3” indicates the seconds
0
≤
m3
≤
59
Query
TM? <terminator>
Example
TM0,0,0
Description • Parameter error 12 will occur if an illegal
value is set.
• While recalling or storing is in progress,
execution error 19 will occur.
• The maximum time that can be specified is
10000 (hours).
WR/WR? Sets the wiring system/inquires about
the current setting.
Syntax
WR m <terminator>
m= 2: 1P3W (for WT230 only)
3: 3P3W (for WT230 only)
4: 3P4W (for WT230 model
760503 only)
5: 3V3A (for WT230 model
760503 only)
Query
WR? <terminator>
Example
WR2
Description Parameter error 12 will occur if an illegal value
is set.
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...